N6754A Datasheet, PDF (22/57 Pages) Keysight Technologies – DC Power Modules
Supplemental Characteristics (continued)
N6751A / N6752A N6753A / N6755A
N6754A / N6756A
Down-programming Settling Time with no load: (Time from start of voltage change to 0.1% of full scale value)
Small voltage step
10 V to 0 V
6 V to 0 V/10 V to 0 V 15 V to 0 V/29 V to 0 V
Time
0.45 ms
0.8 ms/1.3 ms
0.8 ms/1.5 ms
Large voltage step
50 V to 0 V
20 V to 0 V
60 V to 0 V
Time
1.4 ms
2 ms
2.3 ms
Down-programming Time with Capacitive load: (Time from start of voltage change to output voltage < 0.5 V)
Small voltage step
Time
Large voltage step
Time
Capacitive load NOTE 3
10 V to 0 V
2.1 ms
50 V to 0 V
11 ms
1000 µF
6 V to 0 V/10 V to 0 V
2.2 ms/4.5 ms
20 V to 0 V
8.5 ms
4700 µF
15 V to 0 V/29 V to 0 V
2.3 ms/5.5 ms
60 V to 0 V
10 ms
680 µF
Down-programming Capability:
Continuous power
Peak current
7W
12.5 W
12.5 W
7A
15 A
6A
Over-voltage Protection:
Accuracy
Accuracy with Option 761
Accuracy with Option 760
Maximum setting
Response time
0.25% + 0.25 V
0.25% + 0.15V
0.25% + 0.3V
0.25% + 0.25 V
0.25% + 0.45V
0.25% + 0.6V
-
0.25% + 0.45V
0.25% + 0.6V
55 V
22 V
66 V
50 µs from occurrence of over-voltage condition to start of output shutdown
Output Ripple and Noise: (PARD)
CC rms:
2 mA
10 mA
4 mA
Common Mode Noise: (from 20 Hz – 20 MHz; from either output to chassis)
rms
peak-to-peak
500 µA
< 2 mA
500 µA
< 2 mA
750 µA
< 3 mA
Remote Sense Capability:
Outputs can maintain specifications with up to a 1-volt drop per load lead.
The load lead drop reduces the maximum available voltage at the load.
Series and Parallel Operation:
Identically rated outputs can be operated directly in parallel
or can be connected for straight series operation..
Minimum Output Turn-On Delay: (Time from when any Output On command is received until the output starts
turning on)
Without relay option
25 ms
18 ms
18 ms
With relay Option 760
51 ms
44 ms
44 ms
1 Correction On compensates for current flowing into the output capacitor during voltage transients.
2 Stated values apply when the load resistance is at or close to zero. To determine the frequency for larger resistance valuess, refer to the
User’s Guide under “Measurement System Bandwidth”.
3 Modules can discharge the specified capacitive load from full scale to 0V at a rate of 4 times/second.
N6731B-N6786A DC Power Modules
N6700B-N6705B Mainframes
22
Specifications Guide