English
Language : 

N9320BRF Datasheet, PDF (2/11 Pages) Keysight Technologies – RF Spectrum Analyzer
02 | Keysight | N9320B RF Spectrum Analyzer - Data Sheet
Definitions and Conditions
“Specifications” describe the performance of parameters covered by the product
warranty and apply to the full temperature range of 5 to 45 °C, unless otherwise noted.
“Typical” values describe additional product performance information that is not covered
by the product warranty. It is performance beyond specifications that 80 percent of the
units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C.
Typical performance does not include measurement uncertainty.
“Nominal” values indicate expected performance, or describe product performance that
is useful in the application of the product, but are not covered by the product warranty.
The spectrum analyzer
will meet its
specifications when:
It is within its calibration cycle It
has been turned on at least 30
minutes. It has been stored at
an ambient temperature within
the allowed operating range for
at least two hours before being
turned on; if it has been stored
previously at a temperature
range inside the allowed storage
range, but outside the allowed
operating range.