English
Language : 

E1466A Datasheet, PDF (2/4 Pages) Keysight Technologies – 4x64 Relay Matrix Switch
Product Specifications
AC Performance
AC specifications apply with no more than one crosspoint closed per row
or column. Specifications are for 4x64 matrix, for Z(load) = Z(source) =
50 Ω. Specifications are for worst crosspoint. Matrix expansion
degrades crosstalk and bandwidth performance. Typical is defined as the
worst crosspoint test result from one or two matrix modules. If
guaranteed specifications are necessary, contact your local sales
representative.
Crosstalk (dB) within a card (worst path):
<10 kHz <100 kHz <1 MHz
Closed path to closed path (typical):
Open row to open row (typical):
Open row to open column (typical):
Open column to open column (typical):
Crosstalk (dB) module-to-module
(represents 4 x 128 configuration):
–66 dB
–73 dB
–84 dB
–92 dB
<10 kHz
–45 dB
–52 dB
–64 dB
–70 dB
–29 dB
–37 dB
–47 dB
–52 dB
<100 kHz <1 MHz
Closed path to closed path (typical):
Open row to open row (typical):
Open row to open column (typical):
Open column to open column (typical):
–66 dB
–68 dB
–84 dB
–92 dB
–45 dB
–46 dB
–64 dB
–71 dB
–29 dB
–29 dB
–48 dB
–52 dB
Closed channel capacitance (<10 kHz):
Hi to Lo:
<270 pF
Hi to Ground:
<430 pF
Lo to Ground:
<440 pF
Minimum bandwidth
(–3 dB, ZL = ZX = 50 Ω): 10 MHz
Input
Maximum voltage (any terminal to any other terminal or chassis):
DC:
200 V
AC rms:
170 V
Peak:
238 V p-p
Maximum current
(per channel common,
non-inductive):
1 Adc; 1 Aac peak
Maximum power:
Per channel:
30 W
Per module:
62.5 VA (resistive load)
DC
Maximum thermal offset
per channel, differential
Hi-Lo:
5 µV
Closed channel resistance (per channel):
Initial:
<4.0 Ω (worst crosspoint)
<1.8 Ω (best crosspoint)
End of life:
<10.0 Ω
Insulation resistance (between any two points):
≤40 °C, ≤95% RH:
>10E8 Ω
≤40 °C, ≤65% RH:
n/a
≤25 °C, ≤40% RH:
>10E9 Ω
General
Time to close one channel: 8.9 ms (Agilent V/743 and C-SCPI)
Note: When downloading a channel list to card memory, you can close all
columns in one row in 8.9 ms.
Power-down state:
Relay states are unchanged at power-down.
Power-up state:
Relays open at power-up.
Minimum relay life:
No load:
10E7 operations
Screw terminal wire size: 18 to 26 AWG (1.2, 0.9, 0.75, 0.6, 0.5 mm)
2
PDFINFO H 5 5 9 2 - 0 2