|
B1500A Datasheet, PDF (19/36 Pages) Agilent(Hewlett-Packard) – Agilent B1500A Semiconductor Device Analyzer | |||
|
◁ |
19 | Keysight | B1500A Semiconductor Device Analyzer - Data Sheet
Pulse/DC output voltage and accuracy
Output voltage (Vout) 50 ⦠load
â20 V to +20 V
Open load
â40 V to +40 V
Accuracy¹
Open load
±(0.5 % + 50 mV)
Amplitude resolution 50 ⦠load
0.2 mV (|Vout|â¤5V)
0.8 mV (5V<|Vout|â¤20V)
Open load
0.4 mV (|Vout|â¤10V)
1.6 mV (10V<|Vout|â¤40V)
Output connectors
SMA
Source impedance
50 â¦Â²
Short circuit current
800 mA peak (400 mA average³)
Overshoot/
pre-shoot/ringing4
50 ⦠load
±(5 % + 20 mV)
1.â At 1 µs after completing transition.
2.â Supplemental characteristics (±1 %)
3.â This value is specified under the following condition: [(Number of installed
HV-SPGUs) x 0.2 A] + [DC current output by all modules (including HV-SP-
GUs)] < 3.0 A
4.â Follow the specified condition of the transient time.
Pulse range and pulse parameter¹
Frequency range
0.1 Hz to 33 MHz
Pulse period
Programmable range 20 ns to 10 s
Resolution
Minimum4
Accuracy
10 ns
100 ns³
±1 % (±0.01 % + 200 ps)2
Width
Programmable range 10 ns to (period â 10 ns)
Resolution
Minimum4
2.5 ns (Tr and Tf ⤠8 µs)
10 ns (Tr or Tf > 8 µs)
50 ns (25 ns typical)³
Transition time5
(Tr and Tf)
Accuracy
±(3 % + 2 ns)
Programmable range 8 ns to 400 ms
Resolution
Minimum4
2 ns (Tr and Tf ⤠8 µs)
8 ns (Tr or Tf > 8 µs)
15 ns2 (Vamp ⤠10 V)
20 ns (Vamp ⤠10 V)
30 ns (Vamp ⤠20 V)
60 ns (Vamp > 20 V)
Output relay
switching time6
Accuracy
Open/close
â5 % to 5 % + 10 ns (Vamp ⤠10 V)
â5 % to 5 % + 20 ns (Vamp ⤠20 V)
100 µs 2
1.â Unless otherwise stated, all specifications assume a 50 ⦠termination.
2.â Supplemental characteristics.
3.â This is specified at Vamp ⤠10 V.
4.â Minimum value in which timing accuracy can be applied.
5.â The time from 10 % to 90 % of Vamp which is the amplitude of output pulse.
6.â Solid state relay for frequent switching applications.
DPeulfsienwitaivoefnormof pulse waveform
Tperiod
Tdelay
Twidth
Vpeak
Vamp
Vbase
90%
10%
Tr
90%
10%
Tf
SPGU supplemental characteristics
Pulse width jitter: 0.001 % +150 ps
Pulse period jitter: 0.001 % +150 ps
Maximum slew rate: 1000 V/µs (50 ⦠load)
Noise: 10 mVrms (at DC output)
Advanced feature:
Voltage monitor: The HV-SPGU has a voltage monitor
function to measure the voltage at the DUT terminal.
Measurement accuracy (open load): ±(0.1 % of reading +
25 mV)
Measurement resolution: 50 μV
Note: Specified at 1 PLC (20 ms = (5 μs sample + 5 μs
interval) x 2000 samples.)
Voltage compensation: The HV-SPGU can measure the
impedance of DUT and adjust the output voltage according
to the DUT impedance.
ALWG (arbitrary linear waveform generator)
function
Arbitrary linear waveform generator (ALWG) mode:
ââ Output complex waveform per one channel of HV-SPGU
ââ Define multi-level pulse and multi-pulse waveform includ-
ing open state pulse with ALWG GUI editor
ââ Sequential pulse waveform from user-defined pulse waveform
ââ 1024 points per one channel
ââ Programmable timing range: 10 ns to 671.088630 ms,
10 ns resolution
|
▷ |