English
Language : 

N9913A-211 Datasheet, PDF (16/48 Pages) Keysight Technologies – FieldFox Handheld Analyzers
16 | Keysight | FieldFox Handheld Analyzers 4/6.5/9/14/18/26.5/32/44/50 GHz - Data Sheet
Corrected Measurement Uncertainty for N9950A/51A/52A
Full 2-port calibration, 85056D 2.4 mm calibration kit, spec1
Power level of -15 dBm, 10 Hz IF bandwidth, no averaging, battery saver off, and 60-minute warm-up time. Includes uncertainties due
to drift, noise, compression, and dynamic accuracy. Coverage factor of x1 applied to uncertainties, for ease of comparison with other
industry handheld analyzers.
Corrected performance (dB)
Directivity
Source match
Load match
Reflection tracking
Transmission tracking
≤ 2 GHz
42
39
42
± 0.002
± 0.003
Transmission uncertainty (S21, S12)
2 to 20 GHz
34
30
34
± 0.029
± 0.034
20 to 40 GHz
26
23
26
± 0.080
± 0.109
40 to 50 GHz
26
23
26
± 0.075
± 0.105
Reflection uncertainty (S11, S22)
1.  Uncertainty curves shown are calculated based on ISO GUM methodology. The values in the table are provided for reference only, in accordance to legacy
uncertainty methods.