Supplemental Characteristics (continued)
N6773A
N6774A
N6775A
N6776A
N6777A
N6705 Mainframe Oscilloscope Measurement Accuracy: (@ 23 °C ±5 °C; accuracy of any individual point in the trace)
Voltage
0.1% + 45 mV 0.1% + 75 mV 0.1% + 120 mV 0.1% + 160 mV 0.1% + 175 mV
Current
Correction On NOTE 1
0.15% + 45 mA 0.15% + 27 mA 0.15% + 22 mA 0.15% + 12 mA 0.15% + 12 mA
0.15% + 35 mA 0.15% + 22 mA 0.15% + 19 mA 0.15% + 9 mA 0.15% + 9 mA
Up-programming and Down-programming Time with full resistive load:
(Time from 10% to 90% of total voltage excursion; for voltage setting from 0V to full scale and full scale to 0V)
20 ms
20 ms
20 ms
20 ms
20 ms
Maximum Up-programming and Down-programming Settling Time with full resistive load:
(Time from start of voltage change to 0.1% of full-scale value; for voltage setting from 0V to full scale and full scale to 0V)
100 ms
100 ms
100 ms
100 ms
100 ms
Over-voltage Protection:
Accuracy
With Option. 761
With Option. 760
Maximum setting
0.25% +100 mV
0.25% + 500 mV
0.25% + 700 mV
22 V
0.25% + 130 mV
0.25% + 350 mV
0.25% + 700 mV
38.5 V
0.25% + 260 mV
0.25% + 350 mV
0.25% + 400 mV
66 V
0.25% + 650 mV
0.25% + 650 mV
0.25% + 650 mV
110 V
0.25% + 650 mV
0.25% + 650 mV
0.25% + 650 mV
165 V
Response time
50 µs from occurrence of over-voltage condition to start of output shutdown
Output Ripple and Noise (PARD):
CC rms
6 mA
6 mA
6 mA
6 mA
6 mA
Common Mode Noise: (from 20 Hz – 20 MHz; from either output to chassis)
Rms
2 mA
2 mA
2 mA
2 mA
2 mA
Peak-to- peak
< 20 mA
< 20 mA
< 20 mA
< 20 mA
< 20 mA
Remote Sense
Capability:
Outputs can maintain specifications with up to a 1-volt drop per load lead.
The load lead drop reduces the maximum available voltage at the load.
Series and Parallel
Operation:
Identically rated outputs can be operated directly in parallel
or can be connected for straight series operation.
Minimum Output Turn-On Delay: (Time from when any Output On command is received until the output starts turning on)
Without relay
32 ms
32 ms
32 ms
32 ms
32 ms
With Option 760
58 ms
58 ms
58 ms
58 ms
58 ms
1 Correction On compensates for current flowing into the output capacitor during voltage transients.
2 Stated values apply when the load resistance is at or close to zero. To determine the frequency for larger resistance valuess, refer to the
User’s Guide under “Measurement System Bandwidth”.