English
Language : 

DSOX3EMBD Datasheet, PDF (16/26 Pages) Keysight Technologies – Serial Bus Options for InfiniiVision
16 | Keysight | Serial Bus Options for InfiniiVision X-Series Oscilloscopes - Data Sheet
Specifications/Characteristics (Continued)
SENT specifications/characteristics (DSOXT3SENSOR, DSOX4SENSOR, and DSOX6SENSOR)
CAN input source
Analog channels 1, 2, 3, or 4
Digital channels D0 to D15 non-differential
Clock period
1 µs to 300 µs with user-defined tolerance setting from 3 to 30%
Number of nibbles
1 to 6
Idle state
High or low
CRC format
2008 or 2010 standards
Pause pulse
On/Off
Message format
Fast Nibbles (All)
Fast Signals (only)
Fast + Short Serial
Fast + Enhanced Serial (automatically detects bit format: 12-bit data/8-bit ID or 16-bit data/4-bit ID)
Short Serial (only)
Enhanced Serial (only)
Number of defined signals
1 to 6 (each specified by start bit #, number of bits, and nibble order)
Numerical format of signals
Hexadecimal, unsigned decimal, or transfer function with user-defined multiplier and offset for each
defined signal
Triggering
Start of fast channel message
Start of slow channel message
Fast channel status & communication nibble + data
Slow channel message ID
Slow channel message ID + data
Tolerance violation (sync pulse width exceeds user-specified tolerance)
Fast channel CRC error
Slow channel CRC error
All CRC errors
Pulse period error (if nibbles are < 12 or > 27 ticks wide)
Successive sync pulses error (if consecutive sync pulse widths are greater than 1/64 difference)
Fast channel decode
Status & communication nibble (binary digits in green)
Data (hex, unsigned decimal, or transfer function digits in white based on user-defined signal format)
CRC error (hex digit in blue = valid, hex digit in red = error)
Pulse period error (< or > in red)
Slow channel decode
Message ID (hex digits in yellow)
Data (hex digits in white)
CRC (hex digits in blue = valid, hex digits in red = error)
Multi-bus analysis
SENT plus one other serial bus, including another SENT bus