English
Language : 

53181A Datasheet, PDF (16/19 Pages) Keysight Technologies – Counters
16 | Keysight | 53131A/132A/181A Counters - Data Sheet
Measurement Statistics
Available statistics
Number of measurements 2 to 1,000,000
Measurements
Description
Mean, Minimum, Maximum, Standard Deviation
Statistics may be collected on all measurements or on only those which are between the limit bands.
When the limits function is used in conjunction with statistics, N (number of measurements) refers to
the number of in-limit measurements. In general, measurement resolution will improve in proportion
to N, up to the numerical processing limits of the instrument.
Statistics may be collected for all measurements except peak volts and totalize.
General Information
Save and recall
Rack dimensions (HxWxD)
Weight
Warranty
Power supply
AC line selection
Power requirements
Environment
Remote interface
Remote programming language
Safety
EMC
Radiated immunity testing
Description
Up to 20 complete instrument setups may be
saved and recalled later. These setups are
retained when power is removed from the
counter.
88.5 mm x 212.6 mm x 348.3 mm
3.5 kg maximum
3 year standard
100 to 120 VAC ± 10% I 50, 60 or 400 Hz ± 10%
200 to 240 VAC ± 10% I 50 or 60 Hz ± 10%
Automatic
170 VA maximum (30 W typical)
Storage temperature
Operating environment
Operating temperature
Operating humidity
Operating altitude
GPIB (IEEE 488.1-1987 IEEE 488.2-1987)
SCPI-1992.0 (Standard Commands for
Programmable Instruments)
Designed in compliance with IEC-1010,
UL-3111-1 (draft), CAN/CSA 1010.1
Installation Category II
CISPR-11, EN50082-1, IEC 80-2, -3, -4
When the product is operated at maximum
sensitivity (20 mVrms) and tested at 3 V/m
according to IEC 801-3, external 100 to 200 MHz
electric fields may cause frequency miscounts.
—40 °C to 71 °C
Pollution degree 2; indoor locations
0 °C to 55 °C
Up to 80% at 40 °C
Up to 3000 m or 10,000 ft.