34970A/34972A accuracy specifications ± (% of reading + % of range)[1]
Includes measurement error, switching error, and transducer conversion error
Range [3]
Frequency, etc. 24 hour [2]
23 °C ± 1 °C
DC voltage
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
300.000 V
True RMS AC voltage [4]
All ranges from
100.0000 mV
to 100.0000 V
300.0000 V
Resistance [6]
100.0000 Ω
1.000000 kΩ
10.00000 kΩ
100.0000 kΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ
Frequency and period [7]
100 mV
to 300 V
DC current (34901A only)
10.00000 mA
100.0000 mA
1.000000 A
True RMS AC current (34901A only)
10.00000 mA
and [4] 1.000000 A
100.0000 mA [8]
0.0030 + 0.0035
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0020
3 Hz–5 Hz
5 Hz–10 Hz
10 Hz–20 kHz
20 kHz–50 kHz
50 kHz–100 kHz
100 kHz–300 kHz [5]
3 Hz–5 Hz
5 Hz–10 Hz
10 Hz–20 kHz
20 kHz–50 kHz
50 kHz–100 kHz
100 kHz–300 kHz [5]
1.00 + 0.03
0.35 + 0.03
0.04 + 0.03
0.10 + 0.05
0.55 + 0.08
4.00 + 0.50
1.00 + 0.05
0.35 + 0.05
0.04 + 0.05
0.10 + 0.10
0.55 + 0.20
4.00 + 1.25
1 mA current source 0.0030 + 0.0035
1 mA
0.0020 + 0.0006
100 µA
0.0020 + 0.0005
10 µA
0.0020 + 0.0005
5.0 µA
0.002 + 0.001
500 nA
0.015 + 0.001
500 nA || 10 MΩ 0.300 + 0.010
3 Hz–5 Hz
5 Hz–10 Hz
10 Hz–40 Hz
40 Hz–300 kHz
0.10
0.05
0.03
0.006
<0.1 V burden
<0.6 V
<2 V
0.005 + 0.010
0.010 + 0.004
0.050 + 0.006
3 Hz–5 Hz
5 Hz–10 Hz
10 Hz–5 kHz
3 Hz–5 Hz
5 Hz–10 Hz
10 Hz–5 kHz
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.00 + 0.5
0.30 + 0.5
0.10 + 0.5
Temperature
Thermocouple [10]
RTD
Thermistor
Type
B
E
J
K
N
R
S
T
R0 from 49 Ω to 2.1 kΩ
2.2 k, 5 k, 10 k
1-year accuracy [9]
1100 °C to 1820 °C
–150 °C to 1000 °C
–150 °C to 1200 °C
–100 °C to 1200 °C
–100 °C to 1300 °C
300 °C to 1760 °C
400 °C to 1760 °C
–100 °C to 400 °C
–200 °C to 600 °C
–80 °C to 150 °C
1.2 °C
1.0 °C
1.0 °C
1.0 °C
1.0 °C
1.2 °C
1.2 °C
1.0 °C
0.06 °C
0.08 °C
90 Day
23 °C ± 5 °C
1 Year
23 °C ± 5 °C
Temperature
coefficient
0 °C–18 °C, 28 °C–55 °C
0.0040 + 0.0040
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0030
0.0050 + 0.0040
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0030
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0003
1.00 + 0.04
0.35 + 0.04
0.05 + 0.04
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.08
0.35 + 0.08
0.05 + 0.08
0.11 + 0.12
0.60 + 0.20
4.00 + 1.25
1.00 + 0.04
0.35 + 0.04
0.06 + 0.04
0.12 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.08
0.35 + 0.08
0.06 + 0.08
0.12 + 0.12
0.60 + 0.20
4.00 + 1.25
0.100 + 0.004
0.035 + 0.004
0.005 + 0.004
0.011 + 0.005
0.060 + 0.008
0.20 + 0.02
0.100 + 0.008
0.035 + 0.008
0.005 + 0.008
0.011 + 0.012
0.060 + 0.020
0.20 + 0.05
0.008 + 0.004
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.020 + 0.001
0.800 + 0.010
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.040 + 0.001
0.800 + 0.010
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
0.10
0.10
0.05
0.05
0.03
0.03
0.01
0.01
0.005
0.005
0.001
0.001
0.030 + 0.020
0.030 + 0.005
0.080 + 0.010
0.050 + 0.020
0.050 + 0.005
0.100 + 0.010
0.002+ 0.0020
0.002 + 0.0005
0.005 + 0.0010
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.00 + 0.5
0.30 + 0.5
0.10 + 0.5
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.00 + 0.5
0.30 + 0.5
0.10 + 0.5
Extended range 1-year accuracy [9]
400 °C to 1100 °C 1.8 °C
–200 °C to -150 °C 1.5 °C
–210 °C to -150 °C 1.2 °C
–200 °C to -100 °C 1.5 °C
–200 °C to -100 °C 1.5 °C
–50 °C to 300 °C 1.8 °C
–50 °C to 400 °C 1.8 °C
–200 °C to -100 °C 1.5 °C
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006
0.100 + 0.06
0.035 + 0.06
0.015 + 0.06
Temp coefficient/ºC
0.03 °C
0.003 °C
0.002 °C
[1] Specifications are for 1 hr warm-up and 6½ digits, Slow ac filter
[2] Relative to calibration standards
[3] 20% over range on all ranges except 300 Vdc and ac ranges and 1 Adc and
ac current ranges
[4] For sinewave input > 5% of range. For inputs from 1% to 5% of range and
< 50 kHz, add 0.1% of range additional error
[5] Typically 30% of reading error at 1 MHz, limited to 1 x 108 V Hz
16
[6] Specifications are for 4- wire ohms function or 2-wire ohms using scaling to
remove the offset. Without scaling, add 4 Ω additional error in 2-wire Ohms
function
[7] Input > 100 mV. For 10 mV to 100 mV inputs multiply % of reading error x 10
[8] Specified only for inputs >10 mA
[9] For total measurement accuracy, add temperature probe error
[10] Thermocouple specifications not guaranteed when 34907A module
is present. For < 1 °C accuracy, a precision external reference is required.