U1271A Datasheet, PDF (15/22 Pages) List of Unclassifed Manufacturers – U1270 Series Handheld Digital Multimeters | |||
|
◁ |
15 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
Temperature specifications1 - 4
Thermocouple
type
Range
â200 to 1372 °C
K
â328 to 2502 °F
â210 to 1200 °C
J
â346 to 2192 °F
Resolution
0.1 °C
0.1 °F
0.1 °C
0.1 °F
U1271A
1% of reading
+ 1 °C
1% of reading
+ 1.8 °F
â
â
Accuracy ± (% of reading +
as specified below)
U1272A
1% of reading
+ 1 °C
1% of reading
+ 1.8 °F
1% of reading
+ 1 °C
1% of reading
+ 1.8 °F
U1273A/U1273AX
1% of reading
+ 1 °C
1% of reading
+ 1.8 °F
1% of reading
+ 1 °C
1% of reading
+ 1.8 °F
1.â The specifications above is specified after 60 minutes of warm-up time.
2.â The accuracy does not include the tolerance of the thermocouple probe.
3.â Do not allow the temperature sensor to contact a surface that is energized above 30 Vrms or 60 V DC. Such voltages pose a shock hazard.
4.â The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and NIST175.
Capacitance specifications5, 6
Range
10 nF
100 nF
1000 nF
10 μF
100 μF
1000 μF
10 mF
Resolution
0.001 nF
0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
0.001 mF
U1271A
1+5
1+2
1+2
1+2
1+2
1+2
1+2
Accuracy ± (% of reading +
counts of least significant digit)
U1272A
1+5
1+2
1+2
1+2
1+2
1+2
1+2
U1273A/U1273AX
1+5
1+2
1+2
1+2
1+2
1+2
1+2
5.â Overload protection: 1000 Vrms for short circuits with < 0.3 A current.
6.â The accuracy for all ranges is specified based on a film capacitor or better, and after the Null function is used to subtract the test lead resistance and ther-
mal effect (by opening the test leads).
|
▷ |