U2042XA Datasheet, PDF (13/25 Pages) Keysight Technologies – U2040 X-Series Wide Dynamic Range Power Sensors
13 | Keysight | U2040 X-Series Wide Dynamic Range Power Sensors – Data Sheet
General specifications
Inputs / Outputs
Current requirement
Trigger input
Trigger output
Remote programming
Interface
Command language
Maximum measurement speed
Free run trigger measurement
External trigger time-gated measurement
Average mode real time measurement
U2041XA/42XA/43XA/44XA: Approximately < 500 mA
U2049XA: 3W, 802.3af or 802.3at Type 1 standard
Input has TTL compatible logic levels and uses a SMB connector
Output provides TTL compatible logic levels and uses a SMB connector
–– U2041XA/42XA/43XA/44XA: USB 2.0 interface USB-TMC compliance
–– U2049XA: 10/100 Mbps RJ-45 Power Over Ethernet port, transfers data and power on
one single cable, 802.3af or 802.3at Type 1 compliant
SCPI standard interface commands, IVI-COM, IVI-C drivers
25,000 readings per second 1
20,000 readings per second 2
50,000 readings per second 3
1.  Tested under normal mode and fast mode, with buffer mode trigger count of 100, output in binary format,
unit in watt, auto-zeroing, auto-calibration, and step detect disabled.
2.  Tested under normal mode and fast mode, with buffer mode trigger count of 100, pulsed signal with PRF of
20 kHz, and pulse width at 15 µs.
3.  Tested under average only mode and fast mode, with buffer mode trigger count of 200, aperture duration of
20 µs, data format set to real, external trigger or immediate trigger setting.
General characteristics
Environmental compliance
Temperature
Humidity
Altitude
Regulatory compliance
The U2040 X-Series complies with the following
safety and EMC requirements
All models except U2049XA-TVA:
–– Operating condition: 0 to 55 °C
–– Storage condition: –40 to 70 °C
For U2049XA-TVA:
–– Operating Condition: 0 °C to 55 °C
For U2049XA Option TVA, this operating condition is applicable for both standard
atmospheric environment and thermal vacuum environment.
–– Storage condition: –40 °C to 70 °C
–40 °C to 100 °C (for U2049XA Option TVA)
Operating condition: Maximum 95% at 40 °C (non-condensing)
Storage condition: Up to 90% at 65 °C (non-condensing)
Operating condition: Up to 3,000 m (9,840 ft)
Storage condition: Up to 15,420 m (50,000 ft)
IEC 61010-1:2001 / EN61010-1:2001 (2nd edition)
IEC 61326:2002 / EN 61326:1997 + A1:1998 +A3:2003
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS CISPR11:2004
Others
Dimensions (Length x Width x Height)
Weight
Recommended calibration interval
Warranty
U2041XA/42XA/43XA/44XA
168 mm x 46 mm x 35 mm
Net weight: ≤ 0.3 kg
Shipping weight: ≤ 1.3 kg
1 year
3 years
U2049XA
197 mm x 40 mm x 24 mm
Net weight: ≤ 0.37 kg
Shipping weight: ≤ 1.4 kg
1 year
3 years