English
Language : 

N5221A Datasheet, PDF (13/47 Pages) Keysight Technologies – Keysight 2-Port and 4-Port PNA Network Analyzer
Table 4c. N5221A and N5222A with N4433A 4-Port Electronic Calibration Module
Uncertainty curves for the N4433A are created using a 2-port calibration. Multiport uncertainties
are not supported at this time.
Description
Directivity
10 MHz to
50 MHz
50
50 MHz to
500 MHz
50
Specification (dB)
500 MHz
to 2 GHz
50
2 GHz to
13.5 GHz
45
13.5 GHz to
20 GHz
40
Source Match
42
42
42
37
31
Load Match
Reflection Tracking
Mag
Phase (°)
Transmission Tracking
Mag
Phase (°)
40
±0.060
±0.396
±0.068
±0.447
41
±0.060
±0.396
±0.064
±0.421
41
±0.060
±0.396
±0.064
±0.421
35
±0.100
±0.660
±0.115
±0.761
29
±0.180
±1.188
±0.210
±1.387
Transmission Uncertainty, All Options
Reflection Uncertainty, All Options
13