English
Language : 

U3402A Datasheet, PDF (12/21 Pages) Keysight Technologies – U3400 Series 4½ and 5½ Digit Digital Multimeters
12 | Keysight | U3400 Series 4½ and 5½ Digit Digital Multimeters - Data Sheet
Diode test/continuity
U3401A 4.5-digit DMM
Range
2.3000 V
Resolution
100 µV
Maximum reading
2.3000 V
Accuracy (1 year; 23 ºC ± 5 ºC)
0.05% + 5
U3402A 5.5-digit DMM
Rate
Slow
Medium
Fast
Resolution
1.19999 V
2.4999 V
2.499 V
Maximum reading
10.0000 µV
100.00 µV
1.000 mV
Accuracy (1 year; 23 ºC ± 5 ºC)
0.012% + 5
0.012% + 5
0.012% + 2
Frequency
Frequency resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
Range
500.00 Hz
5.0000 kHz
50.000 kHz
500.00 kHz
Measurement range
5 Hz to 500 Hz
500 Hz to 5 kHz
5 kHz to 50 kHz
50 kHz to 500 kHz
Resolution
0.01 Hz
0.1 Hz
1 Hz
10 Hz
Maximum reading
510.00
5.1000
51.000
999.99
Range
500 mV
5V
50 V
500 V
750 V
Input sensitivity for voltage measurement (sine-wave)
5 Hz to 100 kHz
100 kHz to 500 kHz
35 mV RMS
200 mV RMS
0.25 V RMS
0.5 V RMS
2.5 V RMS
5 V RMS
25 V RMS
Not specified
50 V RMS
Not specified
Accuracy (1 year; 23 ºC ± 5 ºC)
0.01% + 5
0.01% + 3
0.01% + 3
0.01% + 3
U3402A 5.5-digit DMM
Range
Measurement range
Resolution
1200.00 Hz
12.0000 kHz
120.000 kHz
1.0000 MHz
5 Hz to 1200 Hz
10 Hz to 12 kHz
100 Hz to 120 kHz
1 kHz to 1 MHz
10 mHz
100 mHz
1 Hz
10 Hz
1.  Specified accuracy at input >5% of full scale.
Maximum reading
1199.99
11.9999
119.999
1.1999
Accuracy (1 year; 23 ºC ± 5 ºC)1
0.005% + 3
0.005% + 2
0.005% + 2
0.005% + 2
Input sensitivity
(sine wave)
40 mV RMS
40 mV RMS
40 mV RMS
0.5 V RMS