English
Language : 

M9391A Datasheet, PDF (12/28 Pages) Keysight Technologies – M9391A PXIe Vector Signal Analyzer
12 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet
Technical Specifications and Characteristics
Dynamic range (continued)
Second harmonic distortion (SHI)
Conversion type: image protect
Pre-amp OFF 25
Frequency
≤ 1.35 GHz
> 1.35 GHz
SHI, nominal26
+35 dBm
+95 dBm
Distortion, nominal27
– 45 dBc
–105 dBc
Nominal Dynamic Range at 2 GHz
-50
Pre-Amp Off, Single - High
-60
-70
-80
-90
-100
-110
-120
-130
-140
-150
-40
-35 -30 -25 -20 -15
Mixer Level (dBm) (Queried)
3rd Order Intermodulation
DANL (1Hz RBW)
-10
-5
DANL (30KHz RBW)
Using 15MHz IF Filters with 60MHz tone spacing for 3rd order intermodulation measurements
Figure 4. Dynamic range at 2 GHz, pre-amp OFF,
single-high conversion type.
Nominal Dynamic Range at 5.8 GHz
-50
Pre-Amp Off, Single - High
-60
-70
-80
-90
-100
-110
-120
-130
-140
-150
-45 -40 -35 -30 -25 -20 -15 -10
Mixer Level (dBm) (Queried)
3rd Order Intermodulation
DANL (1Hz RBW)
DANL (30KHz RBW)
Using 15MHz IF Filters with 60MHz tone spacing for 3rd order intermodulation measurements
Figure 5. Dynamic range at 5.8 GHz, pre-amp OFF,
single-high conversion type.
25. Expected input level -10 dBm. Mixer level offset +10 dB.
26. SHI = second harmonic intercept. The SHI is given by the input power in dBm minus the second harmonic distortion level relative to the input signal in dBc.
27. For 0 dBm input signal.