English
Language : 

N9936A Datasheet, PDF (11/48 Pages) Keysight Technologies – FieldFox Handheld Analyzers
11 | Keysight | FieldFox Handheld Analyzers 4/6.5/9/14/18/26.5/32/44/50 GHz - Data Sheet
Corrected Measurement Uncertainty for N9913A/14A/15A/16A/17A/18A and
N9925A/26A/27A/28A (continued)
Full 2-port calibration, 85518A or 85519A Type-N (m) calibration kit, spec
Power level of -15 dBm, 10 Hz IF bandwidth, no averaging, battery saver off, and 60-minute warm-up time. Includes uncertainties due
to drift, noise, compression, and dynamic accuracy. Coverage factor of x1 applied to uncertainties, for ease of comparison with other
industry handheld analyzers.
Corrected performance (dB)
Directivity
Source match
Load match
Reflection tracking
Transmission tracking
≤ 0.5 GHz
44
37
38
± 0.050
± 0.070
0.5 to 2 GHz
42
36
37
± 0.060
± 0.100
2 to 9 GHz
35
33
31
± 0.070
± 0.180
9 to 18 GHz
32
30
27
± 0.100
± 0.500
Transmission uncertainty (S21, S12)
Reflection uncertainty (S11, S22)