English
Language : 

34951A Datasheet, PDF (11/35 Pages) Keysight Technologies – Multifunction Switch/Measure Unit
11 | Keysight | 34980A Multifunction Switch/Measure Unit - Data Sheet
Table 3. Multiplexer selection table—speciications and characteristics
Channels/conigurations
Switch type
Input characteristics (per channel)
Max volts
Max current (DC, AC RMS)
Switch current
Carry current
Power (W, VA) [6]
Volt-Hertz limit
Initial closed channel res [3][12]
34921A
40 2-wire
20 4-wire
4-current
1.5 A Fused
Armature
latching
± 300 V [1]
1A
2A
60 W
108
< 1.5 Ω
General speciications
Offset voltage [3]
< 3 μV
DC Isolation (ch-ch, ch-earth)
Leakage current [3]
T/C cold junction accuracy [3, 10]
AC characteristics
Bandwidth at terminal block [4]
>10 G Ω
N/A
< 1°C
45 MHz
Crosstalk at terminal block (ch-ch) [4]
300 kHz
–75 dB
1 MHz
20 MHz
45 MHz
Capacitance at terminal block
HI-LO
LO – earth
–75 dB
–50 dB
–40 dB
150 pF
150 pF
34922A
70 2-wire
35 4-wire
Armature
latching
± 300 V [1]
1A
2A
60 W
108
< 1.5 Ω
< 3 μV
>10 GΩ
N/A
N/A
25 MHz
–75 dB
–75 dB
–50 dB
250 pF
200 pF
34923A
80 1-wire
40 2-wire
20 4-wire
Reed
34924A
70 2-wire
35 4-wire
Reed
34925A
80 1-wire
40 2-wire
20 4-wire
Optically
isolated FET
± 150 V peak [2]
0.5 A [5]/ 0.05 A [11]
1.5 A [5]/ 0.05 A [11]
10 W
108
< 1.5 Ω [5]/200 [11]
nominal
± 150 V peak [2]
0.5 A [5]/ 0.05 A [11]
1.5 A [5]/ 0.05 A [11]
10 W
108
< 1.5 Ω [5]/200 [11]
nominal
± 80 V peak [2]
0.02 A [8]
1.6 W
107
< 700 Ω
< 50 μV
< 100 uV 1-wire
>10 GΩ >
N/A
N/A
< 50 uV
10 GΩ
N/A
N/A
< 3 μV
>10 GΩ
20 nA [9]
N/A
45 MHz [5]/4 MHz
10 MHz 1-wire
–75 dB
–75 dB
–50 dB
–40 dB
130 pF
120 pF
25 MHz [5]/4 MHz [11] 1 MHz
–75 dB
–70 dB
–45 dB
Not recommended for
RF signal switching
200 pF
170 pF
100 pF
300 pF (600 pF 1-wire)
[1] DC or AC RMS voltage, channel-to-channel or channel-to-earth
[2] Peak voltage, channel-to-channel or channel-to-earth
[3] Into analog bus. System errors are included in the internal DMM measurement accuracy specifications
[4] 50 Ω source, 50 Ω load, differential measurements verified with 4-port network analyzer (Sdd21)
[5] With input resistors bypassed. Bypassing resistors will reduce lifetime of relays. See the rated load relay life characteristics.
[6] Limited to 6 W of channel resistance power loss per module
[7] Speeds are for 2-wire ohms or DCV, 4 1/2 digits, delay 0, display off, autozero off, and within bank
[8] DC or peak AC current
[9] Ambient temperature < 30°C
[10] Includes 0.5°C temperature reference sensor and 0.5°C terminal block isothermal gradient error, measured under worst case loading of the mainframe; see User’s Guide for
information on supported external reference sensors
[11] With input protection resistors: 2 x 100 Ω ±5%; 0.5W ; TC = ±200 ppm/°C. The series resistance of the 34923/24/25 limits the use of the 100 Ω range.
[12] Channel resistance is typically < 1.5 Ω but can go as high as 50 Ω when a channel is used in measurement applications with < 10 mA load current. Increased relay channel
resistance for measurements with load currents below 10 mA can occur on cards that have been out of service or following relay inactivity for periods of greater than 1 week.
Switching relays for 2K cycles prior to use may reduce the variation in channel resistance. Applies to the 34921A and 34922A. Keysight recommends the use of 4-wire Ohms for
resistance measurements. For high accuracy voltage measurements, select the DMM input resistance setting of >10 G ohms to minimize the impact of relay contact resistance.