English
Language : 

FF-1513-3 Datasheet, PDF (1/1 Pages) JMK Inc. – 1,306,174 Hrs @ 30C GB
The FF-1513-3 filter has been designed to operate at the
input power point of medical equipment.
The filter features low leakage current and the basic filter
is configured to be compatible with FCC and EN55022
requirements.
Agency approvals= UL
Order P/N: FF-1513-3
LINE
BRN
BLUE
GRN/Y
FF-1513-3
LOAD
BRN
BLUE
Operating Voltage (AC)= ______________ 115 / 250 V~
Operating Current= ______________________ 3.0 Amp
Operating Frequency= ___________________ 50/60 Hz
Operating Temperature= _______________-25°to 50°C
Diel. Withstanding (line - case)= ___________ 1500 Vac
Diel. Withstanding (line - line)= ____________ 1500 Vdc
Leakage Current= _____ 60 uA Max. @ 125 VAC, 60 Hz
= ____ 100 uA Max. @ 250 VAC, 50Hz
Max Residual Voltage after 1 sec= __________ 34 Volts
Calcultaed MTBF= ________ 1,306,174 Hrs @ 30°C GB
3.80
3.43
1.75
Ø0.200 (2x)
INSERTION LOSS vs. FREQUENCY
LINE
BROWN
G/Y
3.00
JMK, INC.
FF-1513-3
125/250 VAC
3 AMP
50C
ASSY IN MEXICO
LOAD
BROWN
BLUE
12.0 ±0.5
G/Y
BLUE
12.0 ±0.5
AWG #18 (5PL)
BLUE
USA
EUR
JMK Inc 15 Caldwell Dr. Amherst, NH 03031 PH: 603 886-4100 FX: 603 886-4115
email: info@jmkfilters.com
JMK Inc Glasgow G13 1DN Scotland UK PH: 44-(0) 7785310729 Fax: 44-(0) 141 589 1884