English
Language : 

U1240 Datasheet, PDF (6/9 Pages) Agilent(Hewlett-Packard) – Agilent U1240 Series Handheld Digital Multimeters
Electrical Specifications
Measuring rate
Function
ACV
DCV (V or mV)
Ω
Diode
Capacitance
DCA (μA, mA, A)
ACA (μA, mA, A)
Temperature
Frequency
Times/second
7
7
14
14
4 (< 100 µF)
7
7
7 (single)
1 (> 10 Hz)
General Specifications
Display
Power consumption
Battery type and life
Operating environment
Storage compliance
Safety compliance
Measurement category
EMC compliance
Common Mode Rejection Ratio (CMRR)
Normal Mode Rejection Ration (NMRR)
Crest factor
Temperature coefficient
Shock and vibration
Dimensions (H x W x D)
Weight
Warranty
Dual display (secondary display is intended for temperature function display only)
consists of 4-digit liquid crystal display (LCD) with maximum reading of 11,000 counts.
Automatic polarity indication.
0.22 VA maximum
Four single standard 1.5 V AAA batteries (Alkaline or Zinc Chloride type); 300 hours
typical
• Full accuracy at –10 to 55 °C; and to 80% RH for temperatures up to 30 °C, decreasing
linearly to 50% RH at 55 °C
• 0 – 2000 meters per IEC 61010-1 2nd Edition CAT III, 1000 V / CAT IV, 600 V IEC
61010-1 2nd Edition
–20 to 70 °C
• IEC 61010-1:2001 / EN61010-1:2001
• Canada: CSA C22.2 No. 61010-1:2004
CAT III 1000 V / CAT IV 600 V Overvoltage Protection, Pollution Degree 2
• CISPR 11:1990/EN55011:1990
• Canada: ICES-001:2004
• Australia/New Zealand: AS/NZS CISPR11:2004
> 90 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced)
> 60 dB at 50/60 Hz ± 0.1%
< 3.0
0.1 × (specified accuracy) / °C (from –10 to 18 °C or 28 to 55 °C)
Tested to IEC/EN 60068-2
193.8 mm x 92.2 mm x 58.0 mm
• 450 g with batteries
• 400 g without batteries
• Three years for main unit
• Three months for standard shipped accessories
5