English
Language : 

U1241A Datasheet, PDF (5/6 Pages) Agilent(Hewlett-Packard) – U1241A/U1242A Handheld Digital Multimeter
GENERAL SPECIFICATIONS
Power Supply
4 single standard 1.5 V AAA batteries (Alkaline or Zinc Chloride type)
DISPLAY
Dual display (secondary display is cater for temperature function display only) are 4-digit liquid
crystal display (LCD) with maximum reading of 11,000 counts. Automatic polarity indication.
POWER CONSUMPTION
0.22 VA maximum
OPERATING ENVIRONMENT
Full accuracy at –10 oC to 55 oC
Full accuracy to 80% RH for temperature up to 30 oC, decreasing linearly to 50% RH at 55 oC
STORAGE COMPLIANCE
–20 oC to 70 oC
ALTITUDE
0 – 2000 meters per IEC 61010-1 2nd Edition CAT III, 1000 V
SAFETY COMPLIANCE
• IEC 61010-1:2001 / EN61010-1:2001
• USA: UL 61010-1:2004
• Canada: CSA C22.2 No. 61010-1:2004
MEASUREMENT CATEGORY
CAT III 1000 V Overvoltage Protection, Pollution Degree 2
EMC COMPLIANCE
• Certified to IEC 61326:2002/EN 61326: 2003
• CISPR 11:1990/EN55011:1990
• Canada: ICES-001:2004
• Australia/New Zealand: AS/NZS CISPR11:2004
COMMON MODE REJECTION RATIO (CMRR)
> 90 dB at DC, 50/60 Hz ±0.1% (1kΩ unbalanced)
NORMAL MODE REJECTION RATIO (NMRR)
> 60 dB at 50/60 Hz ±0.1%
CREST FACTOR
< 3.0
TEMPERATURE COEFFICIENT
0.1 × (specified accuracy) / °C (from –10 °C to 18 °C or 28 °C to 55 °C)
SHOCK and VIBRATION
Tested to IEC/EN 60068-2
DIMENSION (HxWxD)
193.8 mm x 92.2 mm x 58.0 mm
WEIGHT
450 g with batteries
400 g without batteries
WARRANTY
3 years
Accessories included:
• Four 1.5 V AAA alkaline batteries
• Certificate of Calibration (CoC)
• Test leads
• Quick Start Guide
• Product Reference CD
Optional Accessories (sold
separately):
• U1162A Alligator clips
• U1163A SMT grabbers
• U1164A Fine tip test probe
• U1181A Immersion probe
(measures food, oil, and other
liquids temperature within the
range of -50 °C to 700 °C)
• U1182A Industrial surface probe
(measures still surface
temperature within the range of
-50 °C and 400 °C)
• U1183A Air probe (measures
dryer, transport pipe and
surrounding air temperature
within the range of -50 °C to
800 °C)
• U1184A Temperature probe
adapter
• U1185A Thermocouple (J-type)
probe and adapter
• U1186A Thermocouple (K-type)
probe and adapter
• U1583A AC current clamp (extend
current measurement range up to
400 A)
DIMENSION
193.8
92
5