U1701B Datasheet, PDF (4/6 Pages) Agilent(Hewlett-Packard) – Agilent U1701B Handheld Capacitance Meter
U1701B Electrical Specifications
Accuracy is given as ± (% of output + counts of least significant digit) at
23 °C ± 5 °C, with relative humidity less than 80% R.H.
For example, 1% ±10 = 1% of reading + 10 counts of least significant digit.
Capacitance
Range
Measuring rate
Resolution Accuracy* (approx.)
1000.0 pF
0.1 pF
1% + 10
5 times/s
10.000 nF
0.001 nF
1% + 5
5 times/s
100.00 nF
0.01 nF
0.5% + 3
5 times/s
1000.0 nF
0.1 nF
0.5% + 3
5 times/s
10.000 μF
0.001μF
0.5% + 3
5 times/s
100.00 μF
0.01 μF
0.5% + 3
5 times/s
1000.0 μF
0.1 μF
0.5% + 3
0.86 times/s
10.000 mF
0.001 mF
1% + 5
0.13 times/s
199.99 mF
0.01 mF
2% + 5
0.006 times/s
* Accuracy is specified to measure film capacitor or better. Use Relative mode to zero residual.
General Specifications
Parameter
Display
Power supply
Power consumption
Battery life
Operating temperature
Operating humidity
Altitude
Storage temperature
Storage humidity
Temperature coefficient
Low battery indicator
Weight
Dimensions (H x W x D)
Safety and EMC
compliance
Calibration
Warranty
U1701B
4½-digit liquid crystal display (LCD) with a maximum resolution of 11,000 counts and automatic polarity
indication
• 9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
• AC power adapter and cord available as options
5.6 mA (on battery operation)
~80 hours without backlight and based on new alkaline
Full accuracy at 0 °C to 50 °C
Up to 80% relative humidity (R.H.) for temperatures up to 31 °C, decreasing linearly to 50% R.H. at 50 °C
0 to 2000 m
–20 °C to 60 °C
0 to 80% R.H. non-condensing
0.1 x (specified accuracy)/ °C (from 0 °C to 18 °C or 28 °C to 50 °C)
will appear when the voltage drops below ~ 6.0 V
320 g
184 mm x 87 mm x 41 mm
IEC 61010-1:2001/EN 61010-1:2001 (2nd Edition) Pollution Degree 2,
IEC 61326-2-1:2005/EN 61326-2-1:2006, ICES-001:2004, AS/NZS CISPR11:2004
One-year calibration cycle recommended
• 3 years
• 3 months for standard shipped accessories
3