English
Language : 

E6651A Datasheet, PDF (4/8 Pages) Agilent(Hewlett-Packard) – Mobile WiMAX test set
E6651A Specifications
Signal analyzer specifications
Signal generator specifications
The test set will meet its warranted performance after one hour within the stated environmental
operating range plus one hour after turn on. Unless otherwise stated all specifications are valid over
the operating temperature range. Supplemental characteristics are intended to provide additional
information, useful in applying the instrument by giving typical (expected), but not warranted,
performance parameters at room temperature (20 to 30 °C). These characteristics are shown in
italics or labeled as nominal.
Frequency
Frequency range
Frequency resolution
Frequency accuracy
Modulation analysis bandwidth
Amplitude
Measurement range1
Absolute measurement accuracy2
Residual response
Spurious responses3
Performance
Quantization
Residual (EVM)4
Pass-band ripple
Triggering
450 MHz to 2.7 GHz (E6651A-503)
450 MHz to 5.99 GHz (E6651A-506)
1 Hz (< 3 GHz), 2 Hz (≥ 3 GHz)
Same as frequency reference
10 MHz
+27 to -50 dBm
±0.5 dB (attenuator 0 to 31 dB)
±1 dB (E6651A-506 attenuator 32 to 62 dB, or
32 to 59 dB for E6651A-503)
-85 dBm at 10 kHz RBW
-55 dBc (excluding IF image)
-40.5 dBc (including IF image) for serial numbers
MY4731xxxx
-45 dBc (including IF image) for serial numbers
MY4815xxxx or later
16-bit ADC (hardware revision A.2.x)
14-bit ADC (hardware revision A.1.x)
≤ 1.5% (symbol)
±0.5 dB (over 16 MHz bandwidth)
Automatic
Frequency
Frequency range
Frequency resolution
Frequency accuracy
Amplitude
WiMAX preamble power range5
CW power range
CW power accuracy6
Resolution
450 MHz to 2.7 GHz (E6651A-503)
450 MHz to 5.99 GHz (E6651A-506)
1 Hz (< 3 GHz), 2 Hz (≥ 3 GHz)
Same as frequency reference
-22 to -102 dBm (< 3 GHz)
-27 to -102 dBm (≥ 3 GHz)
-7 to -127 dBm (< 3 GHz)
-12 to -127 dBm (≥ 3 GHz)
Frequencies < 3 GHz
±0.6 dB (-7 to -78 dBm)
±0.75 dB (≤ -78 to -93 dBm)
±1.0 dB (≤ -93 to -103 dBm)
Frequencies 3 GHz to 5.99 GHz
±1.0 dB (-12 to -95 dBm)
0.01 dB
1 Verified using typical WiMAX signals of 7 MHz, 8.75 MHz and 10 MHz bandwidth
2 Verified using CW measurements. For modulated measurement amplitude accuracy, add +0.5dB to CW amplitude
accuracy.
3 At 2345 MHz, span 16 MHz (for serial numbers MY4731xxxx), span 60 MHz (for serial numbers MY4815xxxx,
MY4832xxxx or MY4914xxxx)
4 Error Vector Magnitude verified using signals of input frequencies from 700 MHz to 5500 MHz, with bandwidths of
7 MHz, 8.75 MHz and 10 MHz and at a level of -20 dBm
5 Verified using typical WiMAX signals of 7 MHz, 8.75 MHz and 10 MHz bandwidth
6 For modulated signal power accuracy, add +0.5dB to CW power accuracy
4