Specifications
Frequency
Range1
Option 520
250 kHz to 20 GHz
Option 532
250 kHz to 31.8 GHz
Option 544
250 kHz to 44 GHz
Resolution
CW
All sweep modes
Switching speed 3, 4
0.001 Hz
0.01 Hz2
< 16 ms (typ) with I/Q modulation off
< 24 ms (typ) with I/Q modulation on
Phase offset
Adjustable in nominal 0.1 ° increments
Frequency bands
Band
Frequency range
N5
1
250 kHz to 250 MHz 1/8
2
> 250 to 500 MHz
1/16
3
> 500 MHz to 1 GHz 1/8
4
> 1 to 2 GHz
1/4
5
> 2 to 3.2 GHz
1/2
6
> 3.2 to 10 GHz
1
7
> 10 to 20 GHz
2
8
> 20 to 28.5 GHz
3
9
> 28.5 to 44 GHz
5
Accuracy
± aging rate ± temperature effects
± line voltage effects (nom)
Internal timebase reference oscillator
Standard8
Aging rate
< ±3 x 10-8/year or
< ±2.5 x 10-10/day
Option UNX
< ±3 x10-8/year or
< ±2.5 x 10-10/day
Temperature effects (typ)
Line voltage effects (typ)
after 30 days
after 30 days
< ±4.5 x 10-9 0 to 55 °C < ±4.5 x 10-9 0 to 55 °C
< ±2 x 10-10 for
< ±2 x 10-10
±10% change
for ±10% change
External reference frequency
10 MHz only
10 MHz only
Lock range
±1.0 ppm
±1.0 ppm
Reference output
Frequency
10 MHz
Amplitude
> +4 dBm into 50 Ω load (typ)
External reference input
Amplitude
Option UNX
5 dBm ±5 dB
5 dBm ±5 dB6
Input impedance
50 Ω (nom)
Step (digital) sweep
Operating modes
Sweep range
Frequency sweep
Amplitude sweep
Dwell time
Number of points
Triggering
Settling time
Frequency
Amplitude
Step sweep of frequency or amplitude or both (start to stop)
List sweep of frequency or amplitude or both (arbitrary list)
Within instrument frequency range
Within attenuator hold range (see “Output” section)
1 ms to 60 s
2 to 65535 (step sweep)
2 to 1601 per table (list sweep)
Auto, external, single, or GPIB
< 8 ms7 (typ)
< 5 ms (typ)
1. Operational, but unspecified, down to 100 kHz.
2. In ramp sweep mode (Option 007), resolution is limited with narrow spans and slow sweep speeds. Refer to ramp sweep
specifications for more information.
3. Time from GPIB trigger to frequency within 0.1 ppm of final frequency above 250 MHz or within 100 Hz below 250 MHz.
4. Add 12 ms (typ) when switching from greater than 3.2 GHz to less than 3.2 GHz.
5. N is a factor used to help define certain specifications within the document.
6. To optimize phase noise 5 dBm ± 2 dB.
7. 19 ms (typ) when stepping from greater than 3.2 GHz to less than 3.2 GHz.
8. Standard performance applies to units with serial numbers ending with 48050000 or greater. For units with lower serial
numbers, refer to the data sheet shipped with the unit or the version of this document dated November 6, 2007.