Table 16. Analyzer environment and dimensions
Description
Supplemental information
General environmental
EMC
Complies with European EMC directive 89/336/EEC, amended
by 93/68/EEC
• IEC/EN 61326
• CISPR Pub 11 Group 1, class A
• AS/NZS CISPR II:2002
• ICES/NMB-001
Safety
Complies with European Low Voltage Directive 73/23/EEC, amended by
93/68/EEC
• IEC/EN 61010-1:2001
• Canada: CSA C22.2 No. 61010-1:2001
• USA: UL 61010-1
Operating environment
Temperature
0 to +40 °C
Instrument powers up and displays no error messages within this
temperature range (except for "source unleveled" error message that
may occur at temperatures outside the specified performance
temperature range of 25 ± 5 °C).
Error-corrected temperature range 23 °C ± 3 °C with less than 1 °C deviation from calibration temp.
Relative humidity
Type-tested 0 to 95% at 40 °C, non-condensing
Altitude
0 to 4600 m (15,000 ft.)
Non-operating storage environment
Temperature
–40 to +70 °C
Cabinet dimensions
Height Width Depth
Excluding front and rear panel hardware and feet
267 mm 426 mm 427 mm
10.5 in 16.75 16.8 in
As shipped - includes front panel connectors,
280 mm 435 mm 470 mm
rear panel bumpers, and feet.
11 in 17.10 in 18.5 in
As shipped plus handles
280 mm 458 mm 501 mm
11 in 18 in 19.7 in
As shipped plus rack-mount flanges
280 mm 483 mm 470 mm
11 in 19 in 18.5 in
As shipped plus handles and rack-mount flanges
280 mm 483 mm 501 mm
11 in 19 in 19.7 in
Weight
Net
N5230A
24.9 kg (55 lb), nominal
Shipping
N5230A
36.3 kg (80 lb), nominal
28