Logic Analyzer Selection Guide for 16900 Series Mainframes
Agilent Model Number
Channels per module
Maximum channels on
single time base
16910A/16911A
102/68
510/340
16950B/16951B
68
340
16760A
34
170
Timing Mode
High-speed timing zoom [1]
Maximum timing sample
rate: half channel mode
Maximum timing sample
rate: full channel mode
Transitional timing
4 GHz (250 ps) with 64 K depth
1.0 GHz (1 ns)
4 GHz (250 ps) with 64 K depth
1.2 GHz (833 ps)
N/A
800 MHz
500 MHz (2.0 ns)
600 MHz (1.67 ns)
800 MHz
500 MHz (2.0 ns)
600 MHz (1.67 ns)
400 MHz
State Mode
Maximum state clock rate
Maximum state data rate
Setup/hold window
Adjustment resolution
State clock, data rate
(upgradeable)
Automated threshold/sample
position, Simultaneous eye
diagrams, all channels
450 MHz with option 500,
250 MHz with option 250
500 Mb/s with option 500,
250 Mb/s with option 250
1.5 ns
80 ps typical
Yes (Agilent E5865A for 16910A)
(Agilent E5866A for 16911A)
Yes
667 MHz
667 Mb/s (DDR)
1066 Mb/s (Dual Sample)
1 ns (600 ps typical),
80 ps typical
No
Yes
800 Mb/s (full channel),
1.5 Gb/s (half channel)
1.5 Gb/s
1 ns
10 ps
No
Yes
Memory Depth [2]
256 M
64 M
32 M
16 M
4M
1M
256 K
Memory depth
(upgradeable)
Option 032
Option 016
Option 004
Option 001
Option 256
Yes (Agilent E5865A for 16910A)
(Agilent E5866A for 16911A)
16951B
16950B, Option 064
16950B, Option 032
16950B, Option 016
16950B, Option 004
16950B, Option 001
Yes (Agilent E5875A)
16760A
64 M standard
Other
Supported signal types
Single-ended
Single-ended and differential
Single-ended and differential
Probe compatibility [3]
40-pin cable connector
90-pin cable connector
90-pin cable connector
Voltage threshold
–5 V to 5 V (10 mV increments) –3 V to 5 V (10 mV increments) –3 V to 5 V (10 mV increments)
Threshold Accuracy
±50 mV + 1% of setting
±30 mV ±2% of setting
±(30 mV + 1% of setting)
[1] All channels, all the time, simultaneous state and timing through same probe.
[2] Specify desired memory depth using available options.
[3] Probes are ordered separately. Please specify probes when ordering to ensure the correct connection between your logic analyzer and the device under test.
2