4294A Datasheet, PDF (1/32 Pages) Agilent(Hewlett-Packard) – Precision Impedance Analyzer
Agilent 4294A
Precision Impedance Analyzer
Data Sheet
Specifications
Specifications describe the instrument’s warranted performance over the tem-
perature range of 0 °C to 40 °C (except as noted). Supplemental performance
characteristics are intended to provide information that is useful in applying
the instrument by given non-warranted performance parameters. These are
denoted as SPC (supplemental performance characteristics), typical, or nomi-
nal. Warm-up time must be greater than or equal to 30 minutes after power on
for all specifications.
Basic Characteristics
Measurement Parameter
Impedance Parameters
|Z|-θ, R-X, Ls-Rs, Ls-Q, Cs-Rs, Cs-Q, Cs-D, |Y|-θ, G-B, Lp-G,
Lp-Q, Cp-G, Cp-Q, Cp-D, Complex Z-Y, |Z|-Ls, |Z|-Cs, |Z|-Lp,
|Z|-Cp, |Z|-Rs, |Z|-Q, |Z|-D, Lp-Rp, Cp-Rp
Measurement Terminal
Configuration
Connector type
Four-terminal pair configuration
Four BNC (female) connectors. Can be converted to one port ter-
minal using the Agilent 42942A Terminal Adapter (7-mm port) or
42941A Impedance Probe (3.5-mm port).
Source Characteristics
Frequency
Range 40 Hz to 110 MHz
Resolution
1 mHz