English
Language : 

4155C Datasheet, PDF (1/14 Pages) Agilent(Hewlett-Packard) – Parameter Analyzer
Agilent 4155C Semiconductor
Parameter Analyzer
Agilent 4156C Precision Semiconductor
Parameter Analyzer
Data Sheet
Introduction
Agilent 4155C and 4156C
Basic functions
• Set measurement and/or stress
conditions
• Control measurement and/or
stress execution
• Perform arithmetic calculations
• Display measured and calculated results
on the LCD display
• Perform graphical analysis
• Store and recall measurement setups, and
measurement and graphical display data
• Dump to printers or plotters
for hardcopy output
• Perform measurement and analysis with
built-in instrument BASIC
• Self test, Auto calibration
Configuration
The 4155C and 4156C are both furnished
with Desktop EasyEXPERT standard edition.
If you want the Desktop EasyEXPERT Plus
edition, you can request the B1541A-002
when you order a 4155C or 4156C. For more
information about the Desktop EasyEXPERT
Plus, please refer to page 11 of this data
sheet.
4155C
4xMPSMU
2xVMU
2xVSU
Desktop
EasyEXPERT
4156C
4xHRSMU
2xVMU
2xVSU
Desktop
EasyEXPERT
41501B (Optional)
2xPGU (Option)¹
GNDU
HPSMU (Option) or
2xMPSMU (Option)¹
1. Minimum number of installable MPSMU
or PGU is two.
SMU: Source monitor unit
Display resolution: 6 digits at each current
range (0.01 fA display resolution at 10
pA range)2
HRSMU: High Resolution SMU
(1 fA/2 μV to 100 mA/100 V)
MPSMU: Medium Power SMU
(10 fA/2 μV to 100 mA/100 V)
HPSMU: High Power SMU
(10 fA/2 μV to 1 A/200 V)
VMU: Voltage Monitor Unit
(0.2 μV resolution in differential mode)
VSU: Voltage Source Unit
PGU: Pulse Generator Unit
(1 channel)
GNDU: Ground Unit
2. Accuracy not guaranteed. Minimum
guaranteed resolution is 1 fA at 10 pA range.