Table 1. Performance Specifications (At 3.3 Vdc supply and 25 C [77 F] unless otherwise noted.)
Parameter
Minimum
Typical
Maximum
Unit
Specific
Note
Interchangeability (first order curve)
0% RH to 10% RH, 90% RH to 100% RH
-7
7
% RH
11% RH to 89% RH
-3
3
% RH
Accuracy (best fit straight line) 11% RH to 89% RH
-3
+3
% RH
4
Hysteresis
2
% RH
Repeatability
0.5
% RH
Settling time
70
ms
Response time (1/e in slow moving air)
5
s
Stability (at 50% RH in 5 years)
1.2
% RH
1
Voltage supply
2.7
5.5
Vdc
2
Current supply
200
500
A
Voltage output (1st order curve fit)
Temperature compensation
VOUT=(VSUPPLY)(0.00636(sensor RH) + 0.1515), typical at 25 C
True RH = (Sensor RH)/(1.0546 0.00216T), T in C
Output voltage temp. coefficient at 50% RH, 3.3 V
-2
mV/ C
Operating temperature
-40[-40] See Figure 2. 85[185]
C[ F]
Operating humidity (HIH-5030)
0
See Figure 2.
100
% RH
3
Operating humidity (HIH-5031)
0
See Figure 2.
100
% RH
Storage temperature
-50[-58]
125[257]
C [ F]
Storage humidity
See Figure 3.
% RH
3
Specific Notes:
1. Includes stress outside of recommended operating zone.
General Notes:
Sensor is ratiometric to supply voltage.
2. Device is tested at 3.3 Vdc and 25 C.
3. Non-condensing environment. When liquid water falls on
Extended exposure to >90 % RH causes a reversible shift
of 3 % RH.
the humidity sensor die, output goes to a low rail condition
indicating no humidity.
Sensor is light sensitive. For best performance, shield
sensor from bright light.
4. Total accuracy including interchangeability is 3 %RH.
2
www.honeywell.com/sensing