HMI63 Datasheet, PDF (1/2 Pages) HITRON ELECTRONICS CORPORTION – UNIVERSAL INPUT AC-DC MEDICAL APPLICATION OPEN FRAME SINGLE OUTPUT 63 WATTS INTERNAL SWITCHING POWER SUPPLIES
H HiTRON
UNIVERSAL INPUT AC-DC MEDICAL APPLICATION OPEN FRAME
SINGLE OUTPUT 63 WATTS INTERNAL SWITCHING
POWER SUPPLIES HMI63 SERIES
FEATURES:
„ ACCOMMODATE UNIVERSAL AC SOURCES
„ COMPACT 4”x2”x1”PCB FORMAT
„ MEET MEDICAL STANDARD EN 60601-1/
UL 2601-1
„ EMI MEET EN 55011 & FCC CLASS B
„ CE MARKING COMPLIANCE
SPECIFICATION
INPUT SPECIFICATION
Input Voltage: Range typ. 90-264Vac.
Rating 100 to 240Vac.
Input Connector: V-M connector.
Input Frequency: 47-63Hz.
Inrush Current: ≤12.0A rms cold start @230Vac.
Input Current: Typ. 0.9A@115Vac, 0.435 @230Vac.
Dielectric Withstand: Meet IEC 60601-1.
EMI: Meet EN 55011 & FCC Class B.
Hold-up Time: Typ. 11.2mS @115Vac, 72mS @ 230Vac.
Earth Leakage: Less than 0.1mA @230Vac.
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Connector: V-M connector.
Output Wattage: Typ. 62 watts.
Line Regulation: Typ. 0.1%.
Load Regulation: Typ. ±1.0%.
Noise & Ripple: Typ. 1.0% peak to peak.
OVP: Built-in at main output (Latch).
Adjustability: Factory Set.
Overload Protection (OLP):
Fully protected against outputs overload and short circuit.
Typ. OLP set at 105-120% of rating output power.
Consult the factory for OLP setting.
GENERAL SPECIFICATION
Efficiency: Typ. 80-89.5 %. (Various with output voltage). Operating Temperature: 0 to +50 °C under forced air flow
Switching Frequency: 85KHz.
for full load. Derate by 1.3W/°C
Circuit Topology: Fixed frequency flyback circuit.
from 40°C to 50°C.
Transient Response: Output voltage returns in less than Storage Temperature: -20 to +85°C.
0.5mS following a 50% load change. Temperature Coefficient: Typ. 0.04%/°C.
Safety Standard: Meet Medical standards UL 2601-1/
Cooling: At least 5 cfm direct forward air flow is required to
EN 60601-1 Class I.
achieve full rating power.
Power Density: 7.7Watts / Cubic Inch.
Construction: 4”x2”x1” PCB format.
Note: Due to requests in market and advances in technology, specifications subject to change without notification.
-15-
For the details of safety approval, please consult the factory.