English
Language : 

HIMP250 Datasheet, PDF (1/2 Pages) HITRON ELECTRONICS CORPORTION – HARMONIC CORRECTION(PFC) AC-DC MEDICAL APPLICATION OPEN FRAME SINGLE OUTPUT 250 WATTS INTERNAL SWITCHING POWER SUPPLIES
H HiTRON
2007/09/10
HARMONIC CORRECTION(PFC) AC-DC MEDICAL APPLICATION
OPEN FRAME SINGLE OUTPUT 250 WATTS INTERNAL SWITCHING
POWER SUPPLIES HIMP250 SERIES
FEATURES:
„ ACCOMMODATE UNIVERSAL AC SOURCES
„ MEET IEC 61000-3-2 HARMONIC
CORRECTION (PFC)
„ MEET MEDICAL STANDARD EN 60601-1/
UL 2601-1
„ EMI MEET EN 55011 & FCC CLASS A
SPECIFICATION
INPUT SPECIFICATION
Input Voltage: Typ. 90-264Vac with PFC.
Input Connector: B-S connector.
Input Frequency: 47-63Hz.
Inrush Current: 11.5Arms @230Vac/50Hz.
Input Current: 2.64A@115Vac, 1.29A @230Vac.
Dielectric Withstand: Meet IEC 60601-1.
EMI: Meet EN 55011 & FCC Class A.
Hold-up Time: Typ. 20.8mS @115Vac, 23.2mS @ 230Vac.
Power Fail Signal: Installed.
Power Factor & Harmonic Correction:
Meet IEC 61000-3-2, PF typ. 0.98-0.99 @ full load.
Over Temperature Protection: By thermostat.
Earth Leakage: Less than 0.0471mA @230Vac.
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Connector: B-S connector.
Output Wattage: Typ. 250 watts.
Line Regulation: Typ. 1.0%.
Load Regulation: Typ. ±0.5%.
Noise & Ripple: Typ. 1.0% peak to peak.
OVP: Built-in.
Adjustability: Available.
Remote on/off: Available.
DC OK Signal: Installed.
Overload Protection (OLP):
Fully protected against outputs overload and short circuit.
Consult the factory for OLP setting.
GENERAL SPECIFICATION
Efficiency: Typ. 83-87 %. (Various with output voltage).
Switching Frequency: 130KHz.
Circuit Topology: PWM fixed frequency forward circuit.
Operating Temperature: 0 to +50 °C under 30 CFM
forced air cooling.
Transient Response: Output voltage returns in less than Storage Temperature: -20 to +85°C.
64mS following a 25% load change.
Safety Standard: Meet Medical standards IEC 60601-1/
UL 2601-1/EN 60601-1 Class I.
Power Density: 3.19Watts / Cubic Inch.
Temperature Coefficient: Typ. 0.04%/°C.
Cooling: 30 CFM air flow is required to deliver the full power.
Construction: U-bracket construction format.
Note: Due to requests in market and advances in technology, specifications subject to change without notification.