English
Language : 

HIMP151_11 Datasheet, PDF (1/2 Pages) HITRON ELECTRONICS CORPORTION – UNIVERSAL INPUT HARMONIC CORRECTION(PFC)
H HiTRON
UNIVERSAL INPUT HARMONIC CORRECTION(PFC)
AC-DC MEDICAL & ITE APPLICATION INTERNAL OPEN FRAME
SWITCHING POWER SUPPLIES 150 WATTS SINGLE, DUAL,
TRIPLE & QUAD OUTPUT HIMP151 SERIES
FEATURES:
 5” X 3” X 1.38” COMPACT SIZE
 ACCOMMODATE UNIVERSAL AC INPUT
 MEET IEC 61000-3-2 HARMONIC
CORRECTION
 EMI MEET CISPR EN 55022/EN 55011 &
FCC CLASS B CONDUCTED
 MEET UNIVERSAL SAFETY STANDARDS
 CE MARKING COMPLIANCE
SPECIFICATION
INPUT SPECIFICATION
Input Voltage: Typical 90-264Vac with PFC.
Input Connector: Molex connector or equivalent.
Input Frequency: 47-63Hz.
Inrush Current: 9.3-11.3Arms at 230Vac.
Input Current: Typical 2.2A at 115Vac, 0.9A at 230Vac.
Dielectric Withstand: Meet IEC 60950-1/IEC60601-1.
EMI: Meet CISPR EN 55022 & FCC Class B conducted.
Meet CISPR EN 55011 & FCC Class B conducted.
Hold-up Time: Typical 90mS at 115Vac & 230Vac.
Power Factor & Harmonic Correction:
Meet IEC 61000-3-2, PF typical 0.99 at full load.
Over Temperature Protection: Installed.
Leakage Current: 0.40-0.50mA at 230Vac touch current.
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Connector: Molex connector or equivalent.
Output Wattage: Max. 150 Watts.
Line Regulation: Typical 1.0%.
Load Regulation: Single O/P VO1 ±3.0%.
Dual O/P VO1 ±3.0% VO2 ±1.0%
Triple O/P VO1 ± 3.0% (± 8% for 3.3V), VO2/VO3 ±3.0%.
Quad O/P VO1/VO2/VO3/VO4 ±3.0%
Noise & Ripple: Typical 1.0-2.0% peak to peak.
OVP: Built-in at VO1 and VO2.
Adjustability: Available at VO1 and VO2.
Overload Protection (OLP):
Fully protected against output overload and short circuit.
OLP set at about 125-150% rating output wattage.
Consult the factory for OLP setting.
GENERAL SPECIFICATION
Efficiency: Typical 76-80%. (various with output)
Switching Frequency: 68K Hz.
Circuit Topology: Single Fixed Frequency Forward circuit.
Dual/Triple/Quad: Fixed Freq. Flyback circuit.
Transient Response: Output voltage returns in less than
14.8-25.2mS following a 25% load change.
Safety Standard: IEC 60950-1 Class I for ITE use.
Operating Temperature: 0 to +50°C under 30 CFM
forced air cooling for full load.
Storage Temperature: -20 to +85°C.
Cooling: Requested 30 CFM/650LFM forced air flow for
full load.
Construction: 4.5”x2.5“x1.2” PCB format.
IEC 60601-1 Class I for Medical use.
Power Density: 6.52-7.24 Watts / Cubic Inch.
NOTE: (1)All measurement are at nominal input, full load and +25°C unless otherwise specifications.
(2)Due to requests in market and advances in technology, specifications subject to change without notification.
-45-