English
Language : 

HIMP151 Datasheet, PDF (1/3 Pages) HITRON ELECTRONICS CORPORTION – UNIVERSAL INPUT HARMONIC CORRECTION AC-DC MEDICAL AND IEC APPLICATION OPEN FRAME SINGLE & MULTI-OUTPUT 150 WATTS INTERNAL SWITCHING POWER SUPPLIES
H HiTRON
2008/02/18
UNIVERSAL INPUT HARMONIC CORRECTION
AC-DC MEDICAL AND IEC APPLICATION OPEN FRAME SINGLE &
MULTI-OUTPUT 150 WATTS INTERNAL SWITCHING POWER
SUPPLIES HIMP151 SERIES
FEATURES:
„ 5” X 3”* X 1.38” COMPACT SIZE
„ ACCOMMODATE UNIVERSAL AC SOURCES
„ MEET IEC 61000-3-2 HARMONIC
CORRECTION
„ MEET UNIVERSAL SAFETY STANDARDS
„ EMI MEET CISPR EN 55022/EN 55011 & FCC
CLASS B CONDUCTED
„ CE MARKING COMPLIANCE
SPECIFICATION
INPUT SPECIFICATION
Input Voltage: Typ. 90-264Vac with PFC.
Input Connector: V-M connector.
Input Frequency: 47-63Hz.
Inrush Current: 9.3-11.3Arms @230Vac.
Input Current: Typ. 2.2A @115Vac, 0.9A @230Vac.
Dielectric Withstand: Meet IEC 60950-1.
EMI: Meet CISPR EN 55022 & FCC Class B conducted.
Meet CISPR EN 55011 & FCC Class B conducted.
Hold-up Time: Typ. 90mS @115Vac & 230Vac.
Power Factor & Harmonic Correction:
Meet IEC 61000-3-2, PF typ. 0.99 @ full load.
Over Temperature Protection: Installed.
Leakage Current: 0.40-0.50mA @230Vac touch current.
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Connector: V-M connector.
Output Wattage: Max. 150 watts.
Line Regulation: Typ. 1.0%.
Load Regulation: Single O/P VO1 ±3.0%.
Dual O/P VO1 ±3.0% VO2 ±1.0%
Triple O/P VO1 ± 3.0% (± 8% for 3.3V), VO2/VO3 ±3.0%.
Quad O/P VO1/VO2/VO3/VO4 ±3.0%
Noise & Ripple: Typ. 1.0-2.0% peak to peak.
OVP: Built-in at VO1 and VO2.
Adjustability: Available at VO1 and VO2.
Overload Protection (OLP):
Fully protected against output overload and short circuit.
OLP set at about 125-150% rating output wattage.
Consult the factory for OLP setting.
GENERAL SPECIFICATION
Efficiency: Typ. 76-80% various with output.
Operating Temperature: 0 to +50°C under 30 CFM
Switching Frequency: 68K Hz.
forced air cooling for full load.
Circuit Topology: Single Fixed frequency forward circuit. Storage Temperature: -20 to +85°C.
Dual/Triple/Quad: Fixed freq. flyback circuit.
Transient Response: Output voltage returns in less than
Temperature Coefficient: Typ. 0.04% /°C.
14.8-25.2mS following a 25% load change. Cooling: Requested 30 CFM forced air flow for full load.
Safety Standard: IEC 60950-1 Class I for IEC use.
Construction: U-bracket format.
IEC 60601-1 Class I for Medical use.
Power Density: 6.52-7.24 Watts / Cubic Inch .
Note: Due to requests in market and advances in technology, specifications subject to change without notification.
For the details of safety approval, please consult the factory.