English
Language : 

L12004-2209H-C Datasheet, PDF (6/8 Pages) Hamamatsu Corporation – Quantum Cascade Laser
PERIPHERAL INSTRUMENTS AND ACCESSORIES
ICables
GHHL Socket Cable A11134-01
A
B
C
No.
A
B
C
Name
HHL TEC drive QCL drive
connector cable
cable
Terminal
9 pin
socket
Cut-off Cut-off
4PS
2PS
(8 wires) (4 wires)
Terminals B and C are to be modified in ac-
cordance with the type of TEC and power
supply.
Pin layout
q
e
r
t
y
u
i
o
!0
Braided outer shield
SP1
SP2
SP3
SP4
White !0
Black q
Red !0
Black q
Green t
Black y
Yellow i
Black o
Braided outer shield
SP5
SP6
White r
Black u
Red r
Black u
GSignal Cable for C11330 A11134-04
Dimensional outline (unit: mm)
D-Sub 9 pin Female
2m
54321
9876
High density D-Sub 15 pin Male
54321
10 9 8 7 6
15 14 13 12 11
Inch screw #4-40
PC Side
D-Sub 9 pin layout
Pin No.
1
2
3
4
5
6
7
8
9
Signal
DCD
RxD
TxD
DTR
GND
DSR
RTS
CTS
RI
Wire for interlock circuit
Milli screw M2.6
Interlock
Frame Ground
C11330 Side
High density D-Sub 15 pin layout
Pin No. Signal Pin No. Signal
1 INTERLOCK 9
GND
2
ALARM
10 Frame Ground
3 RS-422 Rx+ 11
START
4 RS-422 Tx+ 12
STABLE
5 RS-232C Rx 13 RS-422 Rx-
6
GND
14 RS-422 Tx-
7
GND
15 RS-232C Tx
8
GND
IHeatseeker A10767
Heatseeker A10767 consists of 2 types of thermal viewing card and an alignment target. It can
be used for visualization and alignment of the QCL laser beam.
Thermal Viewing Card
Thermal material provides visibility of the IR laser beam. Facilitates tracing of the invisible laser
beam. Two cards with different sensitivity ranges are provided.
Alignment Target
The light axis of the invisible IR laser beam can be easily aligned. Includes a cross target for
checking the light axis. Thermal viewing card can be inserted.
Parameter
Detectable
Thermal viewing card #01
temperature range Thermal viewing card #02
Usable wavelength range
Power required for visibility (*1)
Damage threshold (Max. power density)
Maximum aperture
Storage temperature
Dimensions (W × H × D)
Description / value
Unit
18 to 32
°C
30 to 35
°C
1.0 to 20
µm
>3
mW/mm2
20
mW/mm2
30
mm
-5 to +60 (No condensation) °C
50 × 100 × 12
mm
(*1) Average power density
Dimensional outline (unit: mm)
50
Crossing wire
10
( 0.3)
Laser
beam
2