PHOTOMULTIPLIER TUBE R10699
SPECIFICATIONS
GENERAL
Parameter
Description / value
Unit
Spectral response
185 to 900
nm
Wavelength of maximum response
450
nm
Photocathode
MateriaI
Minimum effective area
Multialkali
8 × 24
—
mm
Window material
UV glass
—
Dynode
Structure
Number of stages
Circular-cage
9
—
—
Direct
Anode to last dynode
4
pF
interelectrode Anode to all other
capacitances electrodes
6
pF
Base
11-pin base JEDEC No. B11-88 —
Weight
Approx. 45
g
Operating ambient temperature
-30 to +50
°C
Storage temperature
-30 to +50
°C
SuitabIe socket
E678–11A (Sold separately)
—
SuitabIe socket assembly
E717–63 (Sold separately)
—
MAXIMUM RATINGS (Absolute maximum values)
Parameter
Value
Unit
Supply Between anode and cathode
1250
V
voltage Between anode and last dynode
250
V
Average anode current A
0.1
mA
CHARACTERISTlCS (at 25 °C)
Parameter
Quantum at 600 nm
efficiency at 780 nm
Cathode
sensitivity
Luminous B
Radiant at 600 nm
at 780 nm
Red / white ratio C
Blue sensitivity index D
Anode
sensitivity
Luminous E
Gain E
Anode dark current F
(Supply voltage at 1 × 106 gain)
Time
response
Anode pulse rise time G
Electron transit time H
Transit time spread (T.T.S.) I
Min.
—
—
620
—
—
—
—
1600
—
—
—
—
—
Typ.
20
10
650
97
63
0.43
15
8500
1.3 × 107
2.0
2.2
22
1.2
Max.
—
—
—
—
—
—
—
—
—
10
—
—
—
Unit
%
%
µA/lm
mA/W
mA/W
—
—
A/lm
—
nA
ns
ns
ns
NOTES
A: Averaged over any interval of 30 seconds maximum.
B: The light source is a tungsten filament lamp operated at a
distribution temperature of 2856 K. Supply voltage is 100 V
between the cathode and all other electrodes connected
together as anode.
C:Red / White ratio is the quotient of the cathode current
measured using a red filter (Toshiba R-68) interposed
between the light source and the tube by the cathode
current measured with the filter removed under the same
conditions as Note B.
D:The value is cathode output current when a blue filter
(Corning CS 5-58 polished to 1/2 stock thickness) is
interposed between the light source and the tube under
the same condition as Note B.
E: Measured with the same light source as Note B and with
the voltage distribution ratio shown in Table 1 below.
Table 1:Voltage distribution ratio
Electrodes K Dy1 Dy2 Dy3 Dy4 Dy5 Dy6 Dy7 Dy8 Dy9 P
Ratio
1 1111 11111
SuppIy Voltage: 1000 V, K: Cathode, Dy: Dynode, P: Anode
F: After 30 min Storage in Darkness
G:The rise time is the time for the output pulse to rise from
10 % to 90 % of the peak amplitude when the entire
photocathode is illuminated by a delta function light pulse.
H:The electron transit time is the interval between the arrival
of delta function light pulse at the entrance window of the
tube and the time when the anode output reaches the
peak amplitube. In measurement, the whole photocathode
is illuminated.
I: Also called transit time jitter. This is the fluctuation in
electron transit time between individual pulses in the signal
photoelectron mode, and is defined as the FWHM of the
frequency distribution of electron transit times