English
Language : 

L10671D_15 Datasheet, PDF (2/4 Pages) Hamamatsu Corporation – S2D2 MODULE
SPECIFICATIONS
S2D2 Lamp (L10671D)
Parameter
Description / Value
Unit
Window Material
UV glass
—
Spectral Distribution
185 to 400
nm
Aperture Diameter
1.0
mm
Output Stability at 230 nm
Drift
Max.
Fluctuation (p-p) Typ.
±0.25
0.005
%/h
%
Guaranteed Life at 230 nm 1
1500
h
Weight
8.6
g
Bulb Wall Temperature 2
Max.
185
°C
Operation Ambient Temperature 3
0 to +40
°C
Storage Temperature
-10 to +60
°C
Operation Ambient Humidity
Below 80 % (no condensation)
—
Storage Humidity
Below 85 % (no condensation)
—
1Life end is defined as the time when the light output intensity at 230 nm falls to 50 % of its initial value or when output fluctuations exceed 0.05 % (p-p).
2Point A in L10671D drawing 3Ambient temperature in open area
Power Supply for S2D2 Lamp (L10671P)
Parameter
Description / Value
Unit
Input Voltage (DC)
8.5 to 13.2
V
Power Consumption
Max.
10
VA
Output Voltage (DC)
With Load
Without Load
Approx. 135
Approx. 250
V
V
Output Current (DC)
30 ± 2
mA
Warm-up Time
25 ± 5
s
D2 Lamp Output Trigger Voltage (DC)
Approx. 250
V
(CN3)
Voltage (DC)
4.2 ± 0.2
V
Filament
Ratings
Warm-up Current (DC)
Time
Operating
Voltage (DC)
Current (DC)
Approx. 0.6
25 ± 5
3.5 ± 0.2
Approx. 0.5
A
s
V
A
CN4 Output
Output Voltage (DC)
Output Current (DC)
Max.
5 ± 0.2
0.15
V
A
CN5 Output 4
Output Voltage (DC) (±)
Output Current (DC) (±)
Max.
5 ± 0.2
0.5
V
A
CN9 Output
Output Voltage (DC)
Output Current (DC)
Max.
12 ± 0.5
0.2
V
A
Weight
42.3
g
Operation Ambient Temperature
0 to +40
°C
Storage Temperature
-10 to +60
°C
Operation Ambient Humidity
Below 80 % (no condensation)
—
Storage Humidity
Below 85 % (no condensation)
—
Conformance Standards
CE Marking
—
Lamp ON / OFF
External Control
CN4 Output ON / OFF
CN5 Output ON / OFF
—
Status signal for lamp ON / OFF, CN4 Output and Main power ON / OFF
4Alternately outputs +5 V and -5 V at 200 msec intervals whenever a signal is input across external control terminals 3 and 10.
Lamp Housing for S2D2 Lamp (L10671H)
Parameter
Description / Value
Unit
Weight
320
g
Optimum Operation Temperature 5
+40 to +80
°C
Storage Temperature
-10 to +60
°C
Operation Ambient Humidity
Below 80 % (no condensation)
—
Storage Humidity
Below 85 % (no condensation)
—
5Point B in L10671H drawing (Before installing this module in equipment, take the heat ratings into account, so the operating temperature will be
within this temperature range.)