English
Language : 

L10101_15 Datasheet, PDF (2/4 Pages) Hamamatsu Corporation – 100 kV MICROFOCUS X-RAY SOURCE
SPECIFICATIONS
GENERAL
Parameter
X-ray tube voltage setting range
X-ray tube current setting range
X-ray tube voltage operational range 1
X-ray tube current operational range 1
Maximum output
X-ray focal spot size (Nominal value) 2
X-ray output window material / Thickness
X-ray beam angle 3
Focus to object distance (FOD)
Target material
Weight 4
Communication method
Description / Value
0 to 100
0 to 200
40 to 100
10 to 200
20
5 to 30
Beryllium/0.15
Approx. 42
Approx. 6.8
Tungsten
Approx. 9.5
Interface: RS-232C (9-pin D-sub connector)
Unit
kV
µA
kV
µA
W
µm
mm
degree
mm
—
kg
—
RATINGS
Parameter
Input voltage (DC)
Power consumption
Rated output
Operating ambient temperature
Storage ambient temperature
Operating and storage humidity
Description / Value
Unit
+24 (+2.4, -0)
V
Less than 120
W
Continuous rating
—
+10 to +40
°C
0 to +50
°C
20 to 85 (No condensation)
%
REGULATION AND STANDARD
Parameter
RoHS directive
EMC
Description
Unit
EN 50581 Category 9
—
IEC/EN 61326-1
Emission limits: CISPR 11 Group 1 Class A
Immunity requirements: Table 2
—
CONTROL SOFTWARE 5
Parameter
Description
Applicable PC
Applicable OS
PC / AT compatible
Windows® XP, 7
Interface
RS-232C interface
NOTE: 1See the graph of the X-ray tube Voltage / Current operation range.
2This focal spot size changes depending on the output.
3Reference value: With 50 % of maximum X-ray emission.
4This weight includes the accessories of approx. 0.25 kg.
5The control software is provided as a sample software for the purpose of MFX operation.
The performance of the software is not guaranteed.
Unit
—
—
—
! PRECAUTIONS TO USE
· This microfocus X-ray source generates X-rays harmful to the human body. Use sufficient caution when handling the equipment to avoid
direct or inadvertent exposure to X-rays.
Install the X-ray source or the X-ray tube unit in an X-ray shielded cabinet or room equipped with safety interlock functions to prevent
accidental exposure to X-rays.
OPERATIONAL CAUTION
· This microfocus X-ray source generates X-rays and must therefore be used only under the supervision of qualified personnel.
· This microfocus X-ray source shall be used in compliance with health and safety regulations enforced in order to prevent health hazards
problems due to ionizing radiation.