English
Language : 

C9692 Datasheet, PDF (2/8 Pages) Hamamatsu Corporation – PHOTON DETECTION UNIT
SPECIFICATIONS
C9692 SERIES COMMON SPECIFICATIONS
Parameter
Description / Value
Unit
Detection Method
Photon counting method
—
Spectral Response Range
Photocathode Size
Maximum Count Rate
185 to 650
nm
16 × 18
mm
3 × 106
s-1
Counter Gate Time
Dark Count (Typ. at 25 °C)
Counter Capacity
0.001 to 10 (1, 2, 5 Steps)
s
50
s-1
32 bits/gate
—
Trigger Signal Input Mode
External trigger, software trigger
—
Trigger Section Trigger Signal Level
TTL negative logic
—
Trigger Signal Pulse Width
100 ns or more
—
Input Voltage (DC)
+7 (supplied from AC adapter)
V
Input Voltage (AC) to Supplied AC Adapter
100 V to 240 V (auto switchable), single phase 50 Hz/60 Hz
—
Operating
Temperature
Humidity
+5 to +40
°C
Below 80 (no condensation)
%
Storage
PC and
Recommended
System
Requirements
Temperature
Humidity
CPU
CLK
Memory
OS
Interface
0 to +50
°C
Below 85 (no condensation)
%
Intel® Pentium® 3 or higher
—
1 GHz or higher
—
512 MB or more
—
Windows® 98, 98SE, Me, 2000, XP-Pro
—
USB Ver1.1
—
C9692-01
Parameter
Optical Fiber Adapter
Distance to Photocathode
Weight
Description / Value
Unit
FC type (HRFC-R1/Hirose)
—
26.5 (from fiber end)
mm
Approx. 0.7
kg
C9692-02
Parameter
Description / Value
Unit
Effective Size of Sample Compartment (W × D × H)
50 × 50 × 15
mm
Distance to Photocathode
45.5 (from bottom of sample compartment)
mm
Weight
Approx. 1.1
kg
C9692-03
Parameter
Description / Value
Unit
Excitation Light
Source (UV LED)
Wavelength · Output Power
Irradiation Time
Irradiation Area
375 nm · 10 mW/cm2
0.1 to 3600
10 dia. (center of sample compartment)
—
sec.
mm
Effective Size of Sample Compartment (W × D × H)
50 × 50 × 15
mm
Distance to Photocathode
62.5 (from bottom of sample compartment)
mm
Weight
Approx. 1.2
kg