English
Language : 

C9407MA Datasheet, PDF (2/5 Pages) Hamamatsu Corporation – Compact and low cost, OEM model (C9409MA)
Mini-spectrometer RC series C9407MA, C9409MA
s Optical characteristics
Parameter
RC-VIS-MOS
C9407MA (Spectrometer module)
RC-VIS-MOS
OEM model C 9 4 09 M A (Spectro m eter head)
Unit
Spectral response range
340 to 780
nm
Spectral resolution Max.
(Spectral response half width)*2
9
nm
Wavelength reproducibility *3
±0.5
nm
Wavelength temperature dependence
0.05
nm/°C
Spectral stray light *2 *4
-30
dB
Broadband stray light *2 *5
-25
dB
*2: Depends on the slit opening. Values were measured with the slit listed in the table “s General ratings / Absolute maximum
ratings”.
*3: Measured under constant light input conditions.
*4: When light at 550 nm is input, spectral stray light is defined as the ratio of the count measured at the input wavelength, to the
count measured in a region of the input wavelength ±40 nm.
*5: The ratio of the transmittance in the transmitting wavelength region of an optical filter (OG530) to that in the blocking region.
s Electrical characteristics
Parameter
A/D conversion
Integration time
Interface
C9407MA (Spectrometer module)
16
5 to 10000
USB1.1
OEMmodel C 9 4 09 M A (Spectro m eter head) Unit
-
bit
-
ms
-
-
s General ratings/Absolute maximum ratings
Parameter
C9407MA (Spectrometer module) OEMmodel C 9 4 09 M A (Spectro m eter head)
Dimensions
73 (W) × 53 (D) × 46 (H)
28 (W) × 28 (D) × 28 (H)
Built-in head
C9409MA
-
Image sensor
CMOS linear image sensor (S8378-256N)
Number of pixels
256
Slit *6
70 (H) × 550 (V)
Optical NA
0.22
Fiber core diameter
600
Optical fiber connector
SMA905
Operating temperature *7
+5 to +40
Storage temperature
-20 to +70
*6: Entrance slit aperture size of the incorporated image sensor.
*7: No condensation
Unit
mm
-
-
pixels
µm
-
µm
-
°C
°C
s Dimensional outlines (unit: mm)
C9407MA
(4 ×) M2.6 DEPTH 3
C9409MA
C9409MA
C9409MA
(2 ×) M2.6 DEPTH 5
FLEXIBLE BOARD
11.5
50.0
73.0
340
14
46
28
320
KACCA0147EC
28
KACCA0149EC
2