English
Language : 

S8982 Datasheet, PDF (1/7 Pages) Hamamatsu Corporation – CCD area image sensor Front-illuminated FFT-CCDs for X-ray imaging
IMAGE SENSOR
CCD area image sensor
S8982, S10127
Front-illuminated FFT-CCDs for X-ray imaging
S8982 is an FFT-CCD image sensor ideal for intra-oral X-ray imaging in dental diagnosis. S8982 has 1.5 mega (1500 × 1000) pixels each of
which is 20 × 20 µm. S8982 delivers low dark current when operated in MPP (Multi Pinned-Phase) mode.
S10127 is an easy-to-use X-ray imaging module using S8982, with added functions such as a cable assembly and X-ray trigger circuit. S10127 is
low cost type for S8981-02 and adapted both AC and DC X-ray sources.
Features
l Sensor has X-ray monitoring photodiode
l Compactness
5.1 mm thickness excluding I/O connector part
l High dynamic range: 12 bit
l Long-term stability
For use under 100,000 shots
(60 kVp, 30 mR X-ray irradiation)
l Resolution: 10 Lp/mm
l 1500 (H) × 1000 (V) pixel format
l Pixel size: 20 × 20 µm
l Coupled with GOS for X-ray imaging
l 100 % fill factor
l Low dark signal
l Low readout noise
l MPP operation
l AC/DC X-ray source adapted
Applications
l General X-ray imaging
l Non-destructive inspection
l Intra-oral X-ray imaging in dental diagnosis
s Selection guide
Type No.
S8982
S10127
s General ratings
Parameter
CCD structure
Fill factor
Number of active pixels
Pixel size
Active area
Vertical clock phase
Horizontal clock phase
Output circuit
Dimensional outline
Reliability
Window
Other
Cooling
Non-cooled
Number of
total pixels
1508 × 1002
Number of
active pixels
1500 × 1000
Active area
[mm (H) × mm (V)]
30 × 20
S8982
S10127
Full frame transfer
100 %
1500 (H) × 1000 (V)
20 (H) × 20 (V) µm
30 (H) × 20 (V) mm
2 phase
2 phase
Emitter follower without load resistance
35.5 (H) × 23.2 (V) mm
38.7 (H) × 26.3 (V) mm
100,000 shots at 60 kVp, 30 m Roentgen
GOS film
MPP mode (low dark current operation), module (S10127)
1