S10814 Datasheet, PDF (1/10 Pages) Hamamatsu Corporation – Front-illuminated FFT-CCD for X-ray imaging
CCD area image sensors
S10814
S10811
Front-illuminated FFT-CCD for X-ray imaging
The S10814 is an FFT-CCD image sensor suitable for intra-oral X-ray imaging in dental diagnosis. The S10814 has about 2 mega
(1700 × 1200) pixels, each of which is 20 × 20 μm in size. The FOP (fiber optic plate) used as an input window is as thin as 1.5
mm, making high resolution as well as highly resistant to X-ray irradiation. The scintillator coated on the FOP is optimized to have
high X-ray sensitivity and high resolution (20 Lp/mm).
The S10811 is an easy-to-use X-ray imaging module using the S10814, with added functions such as a cable assembly and X-ray
trigger circuit.
Features
X-ray monitoring photodiode incorporated
Compact size
High dynamic range: 12-bit
Long-term stability
For use under 100000 shots
(60 kVp, 30 mR X-ray irradiation)
Resolution: 20 Lp/mm
1700 (H) × 1200 (V) pixel format
Pixel size: 20 × 20 μm
Coupled with FOS for X-ray imaging
100% fill factor
Low dark signal
Low readout noise
MPP operation
AC/DC X-ray source adapted
Applications
Intra-oral X-ray imaging in dental diagnosis
General X-ray imaging
Non-destructive inspection
These products are components for incorporation into medical
device.
General ratings
Parameter
CCD structure
Fill factor
Cooling
Number of pixels
Number of active pixels
Pixel size
Active area
Vertical clock phase
Horizontal clock phase
Output circuit
Dimensions
Reliability
Window
S10814
S10811
Full frame transfer
100 %
Non-cooled
1708 (H) × 1202 (V)
1700 (H) × 1200 (V)
20 (H) × 20 (V) μm
34 (H) × 24 (V) mm
2 phases
2 phases
Emitter follower without load resistance
37.8 (H) × 26.5 (V) mm
42.3 (H) × 30.0 (V) mm
100000 shots at 60 kVp, 30 mR
FOS (scintillator on 1.5 mm FOP)
www.hamamatsu.com