L7810_09 Datasheet, PDF (1/4 Pages) Hamamatsu Corporation – Light level control of light source Sensitivity control of optical sensor
CALIBRATED LAMP
LIGHT LIGHT SOURCE SERIES
SOURCE L7810, L7810-02 (Xenon Lamp)
L7820, L7820-02 (Deuterium Lamp)
Easy to reproduce highly stable light!!
L7810, L7810-02
Left: Lamp Housing
Right: Power Supply
L7820, L7820-02
TLSOF0151
Left: Lamp Housing
Right: Power Supply
These light sources deliver the extremely high levels of "stability" and
"repeatability" essential to calibrated light sources. These are availa-
ble as an optimal set including a lamp, lamp housing and power sup-
ply, so that anyone can easily reproduce a highly stable light output.
The L7810-02 xenon lamp light source is calibrated over a wide spec-
tral range from 200 nm to 800 nm, while the L7820-02 deuterium lamp
light source is calibrated in the UV range from 200 nm to 400 nm.
Since we are certified as an ASNITE-Calibration laboratory, we are
capable of calibrating spectral irradiance over a wide range from 200
nm to 800 nm (L7810-02).
The spectral range at wavelengths shorter than 250 nm is calibrated
based on the ASNITE calibration system (Accreditation System of Na-
tional Institute of Technology and Evaluation), while the spectral
range at wavelengths longer than 250 nm is calibrated based on the
JCSS (Japan Calibration Service System).
As of August 2009, Hamamatsu Photonics is the only company certi-
fied as an ASNITE-Calibration laboratory in the "Light" field.
Manufactured upon receiving your order.
TLSOF0159
APPLICATIONS
GLight level control of light source
GSensitivity control of optical sensor
GLight intensity measurement
and studies of photoreactions
(Light resistance, light curing, etc.)
GQuality control of photometric
equipment
(Medical analysis equipment, semi-
conductor inspection systems,
imaging devices, etc)