English
Language : 

H9319_15 Datasheet, PDF (1/2 Pages) Hamamatsu Corporation – Head-on PMT
Head-on PMT
Photon Counting Head H9319 Series
The H9319 series photon counting heads are designed to perform photon
counting by just connecting to a PC. The H9319 series includes a 25-mm (1")
diameter head-on photomultiplier tube, a photon counting circuit, a high-volt-
age power supply circuit, counter and a microprocessor. Data transfer, meas-
urement time and other necessary adjustments can be controlled by com-
mands from the PC through the RS-232C interface.
Since the H9319 series performs linearity correction by the internal micro-
processor, it provides excellent count linearity within a range of ±1 % at
20 × 106 s-1.
Product Variations
Type No. Spectral Response
Sample Program
H9319-01
yes
300 nm to 650 nm
H9319-11
no
H9319-02
yes
300 nm to 850 nm
H9319-12
no
This product can't be used at vacuum environment or reduced pressure environment.
Prescaler
1/4
1/4
1/4
1/4
Specifications
Parameter
H9319-01
H9319-11
H9319-02
H9319-12
Input Voltage
+4.75 to +5.25
Max. Input Voltage
+6
Max. Input Current
60
Effective Area
22
Peak Sensitivity Wavelength
420
300 nm
2.3 × 105
2.1 × 105
Count
Typ.
Sensitivity
400 nm
500 nm
600 nm
4.1 × 105
3.4 × 105
5.7 × 104
2.5 × 105
2.0 × 105
1.3 × 105
700 nm
—
7.8 × 104
Count Linearity *1
20 × 106
Typ.
150
Dark Count *2
Max.
300
10 000
15 000
PMT Operating Voltage Range
+300 to +1200
Integration Time
10 to 1000
1 *3
Settling Time
5 *4
Input Signal (External Trigger Input) *5
TTL level signal
Output Signal (User Line Output) *6
TTL level signal
Interface
RS-232C, 9600 baud, Parity none, 8 data bit, 1 stop bit
Operating Ambient Temperature *7
+5 to +50
Storage Temperature *7
-20 to +50
Weight *8
280
*1: Random pulse, within ±1 % count loss (by count linearity compensation)
*2: After 30 minutes storage in darkness
*3: The time required for the output to reach a stable level following a change in the control voltage from 500 V to 1000 V in darkness
*4: The time required for the output to reach a stable level following a change in the control voltage from 1000 V to 500 V in darkness
*5: Selectable ECR mode or LCR mode (refer to command list)
*6: Controllable by RS-232C command
*7: No condensation
*8: Main body: Approx. 180 g
(at +25 °C)
Unit
V
V
mA
mm
nm
s-1·pW-1
s-1
s-1
V
ms
s
s
—
—
—
°C
°C
g
64