H8249 Series
The H8249 series photosensor modules incorporate a 28-mm (1-1/8")
diameter side-on photomultiplier tube, a high-voltage power supply circuit and
a low noise amplifier. Two types of amplifiers are available with a current-to-
voltage conversion factor of 1 V/µA or 0.1 V/µA and a frequency bandwidth of
DC to 20 kHz or DC to 200 kHz. Three types of photomultiplier tubes are
provided as standard lineups for general applications in the visible range, UV
to visible range requiring low-noise, and UV to near IR range.
Product Variations
Type No.
H8249-001
H8249-011
H8249-101
H8249-002
H8249-012
H8249-102
Spectral Response
Current-to-Voltage
Conversion Factor
Frequency Bandwidth
Features
300 nm to 650 nm
For general applications in visible range
185 nm to 680 nm
185 nm to 900 nm
1 V/µA
DC to 20 kHz
For general applications in UV to visible range
High sensitivity in UV to near IR range. Uses photomultiplier
tube with meshless grid for excellent uniformity.
300 nm to 650 nm
For general applications in visible range
185 nm to 680 nm
185 nm to 900 nm
0.1 V/µA
DC to 200 kHz
For general applications in UV to visible range
High sensitivity in UV to near IR range. Uses photomultiplier
tube with meshless grid for excellent uniformity.
Specifications
Parameter
H8249 Series
Unit
Suffix
-001 / -002
-011 / -012
-101 / -102
—
Input Voltage
±11.5 to ±15.5
V
Max. Input Voltage
±18
V
Max. Input Current
+45/-1 (-001/-011/-101), +55/-15 (-002/-012/-102)
mA
Max. Control Voltage
+1.2 (Input Impedance 100 kΩ)
V
Recommended Control Voltage Adjustment Range
+0.3 to +1.1
V
Effective Area
4 × 20
mm
Sensitivity Adjustment Range
1: 104
—
Peak Sensitivity Wavelength
400
nm
Luminous Sensitivity
Min.
Typ.
30
60
40
60
140
250
µA/lm
Blue Sensitivity Index (CS 5-58) Typ.
7.1
6.4
8.0
—
Red/White Ratio
Typ.
—
—
0.3
—
Radiant Sensitivity *1 Typ.
60
60
74
mA/W
Suffix (with internal 20 kHz amp)
-001
-011
-101
Luminous Sensitivity *2
Min.
Typ.
5.0 × 107
6.0 × 108
2.0 × 108
4.0 × 108
4.0 × 108
2.5 × 109
V/lm
Radiant Sensitivity *1 *2 Typ.
600
400
740
V/nW
Voltage Output Depending Typ.
5
on PMT Dark Current *2 *3 *4 Max.
50
0.1
2
3
50
mV
Max. Output Signal Voltage
+10 (Load resistance 10 kΩ)
V
Current-to-Voltage Conversion Factor
1
V/µA
Suffix (with internal 200 kHz amp)
-002
-012
-102
Luminous Sensitivity *2
Min.
Typ.
5.0 × 106
6.0 × 107
2.0 × 107
4.0 × 107
4.0 × 107
2.5 × 108
V/lm
Radiant Sensitivity *1 *2 Typ.
60
40
74
V/nW
Voltage Output Depending Typ.
0.5
on PMT Dark Current *2 *3 *4 Max.
5
0.01
0.2
0.3
5
mV
Max. Output Signal Voltage
+10 (Load resistance 10 kΩ)
V
Current-to-Voltage Conversion Factor
0.1
V/µA
H8249 series
Offset Voltage
Typ.
±3
mV
Ripple Noise *2 *5 (peak to peak) Max.
0.6
mV
Settling Time *6
Max.
0.2
s
Operating Ambient Temperature *7
+5 to +45
°C
Storage Temperature *7
-20 to +50
°C
Weight
220
g
*1: Measured at the peak sensitivity wavelength *2: Control voltage = +1.0 V *3: After 30 minutes storage in darkness
*4: Output of anode dark current *5: Cable RG-174/U, Cable length 450 mm, Load resistance = 1 MΩ, Load capacitance = 22 pF
*6: The time required for the output to reach a stable level following a change in the control voltage from +1.0 V to +0.5 V. *7: No condensation