English
Language : 

H7827 Datasheet, PDF (1/2 Pages) Hamamatsu Corporation – Compact Head-on PMT
Compact Head-on PMT
Photosensor Modules H7827 Series
The H7827 series photosensor modules incorporate a 19-mm (3/4") diameter
head-on photomultiplier tube, a high-voltage power supply circuit and a low
noise amplifier. Two types of amplifiers are available with a current-to-voltage
conversion factor of 1 V/µA or 0.1 V/µA and a frequency bandwidth of DC to
20 kHz or DC to 200 kHz. Two types of photomultiplier tubes with different
spectral response characteristics are provided for measurement in the visible
range or visible to near IR range.
Product Variations
Type No.
H7827-001
H7827-011
H7827-002
H7827-012
Spectral Response Current-to-Voltage Conversion Factor Frequency Bandwidth
Features
300 nm to 650 nm
300 nm to 850 nm
1 V/µA
DC to 20 kHz
For general applications in visible range
For visible to near IR range
300 nm to 650 nm
300 nm to 850 nm
0.1 V/µA
DC to 200 kHz
For general applications in visible range
For visible to near IR range
Specifications
Parameter
H7827 Series
Suffix
Input Voltage
Max. Input Voltage
Max. Input Current
Max. Control Voltage
-001 / -002
-011 / -012
±11.5 to ±15.5
±18
+40/-1
+1.2 (Input Impedance 100 kΩ)
Recommended Control Voltage Adjustment Range
+0.5 to +1.1
Effective Area
Sensitivity Adjustment Range
Peak Sensitivity Wavelength
Luminous Sensitivity
Min.
Typ.
Blue Sensitivity Index (CS 5-58)
Red/White Ratio
Radiant Sensitivity *1
15
1: 103
420
380
60
80
90
120
10.5
—
—
0.2
85
49
Suffix (with internal 20 kHz amp)
Luminous Sensitivity *2
Min.
Typ.
Radiant Sensitivity *1 *2
Voltage Output Depending Typ.
on PMT Dark Current *2 *3 Max.
Max. Output Signal Voltage
Current-to-Voltage Conversion Factor
-001
1.0 × 107
5.0 × 107
47
3
20
+10 (Load resistance 10 kΩ)
1
-011
1.0 × 107
3.0 × 107
13
3
20
Suffix (with internal 200 kHz amp)
Luminous Sensitivity *2
Min.
Typ.
Radiant Sensitivity *1 *2
Voltage Output Depending Typ.
on PMT Dark Current *2 *3 Max.
Max. Output Signal Voltage
-002
1.0 × 106
5.0 × 106
4.7
0.3
2
+10 (Load resistance 10 kΩ)
-012
1.0 × 106
3.0 × 106
1.3
0.3
2
Current-to-Voltage Conversion Factor
0.1
Offset Voltage *2
Typ.
Ripple Noise *2 *4 (peak to peak) Max.
Settling Time *5
H7827 series
±3
4
0.2
Operating Ambient Temperature
+5 to +45
Storage Temperature
Weight
-20 to +50
80
*1: Measured at the peak sensitivity wavelength *2: Control voltage = +1.0 V *3: After 30 minute storage in darkness
*4: Cable RG-174/U, Cable length 450 mm, Load resistance = 1 MΩ, Load capacitance = 22 pF
*5: The time required for the output to reach a stable level following a change in the control voltage from +1.0 V to +0.5 V.
34
Unit
—
V
V
mA
V
V
mm
—
nm
µA/lm
—
—
mA/W
V/lm
V/nW
mV
V
V/µA
V/lm
V/nW
mV
V
V/µA
mV
mV
s
°C
°C
g