English
Language : 

H7732_08 Datasheet, PDF (1/2 Pages) Hamamatsu Corporation – photosensor modules consist of a 28-mm (1-1/8") diameter side-on photomultiplier tube
Photosensor Modules
H7732 Series
Product Variations
Type No.
H7732
H7732-01
Spectral Response
300 nm to 650 nm
185 nm to 680 nm
H7732-10
185 nm to 900 nm
H7732P-01
H7732P-11
185 nm to 680 nm
185 nm to 850 nm
The H7732 series photosensor modules consist of a 28-mm (1-1/8") diameter
side-on photomultiplier tube and a high-voltage power supply. These side-on
photomultiplier tubes have long been used for spectroscopic applications and
provide high gain and high sensitivity. Five types of photomultiplier tubes are
provided as standard lineups to meet various needs for spectral response range.
Connectors are used for power input and signal output. By selecting cables in
convenient length, the H7732 can be easily installed inside equipment or re-
moved from equipment.
The H7732 is a general-purpose type and the H7732-01 is a low-noise type.
The H7732-10 is sensitive over a wide range from UV to near infrared and
has particularly high sensitivity in wavelengths above 600 nm. The H7732P-
01 and H7732P-11 are selected as low dark count types ideal for photon
counting and low-light-level measurement.
Features
For general applications in visible range
Low noise in UV to visible range
High sensitivity in UV to near IR range.
Uses photomultiplier tube with meshless grid for excellent uniformity.
For photon counting
Specifications
Parameter
H7732
H7732-01 H7732-10 H7732P-01 H7732P-11
Unit
Input Voltage
+11.5 to +15.5
V
Max. Input Voltage
+18
V
Max. Input Current
40
mA
Max. Output Signal Current
100
µA
Max. Control Voltage
+1.2 (Input impedance 100 kΩ)
V
Recommended Control Voltage Adjustment Range
+0.3 to +1.1
V
Effective Area
4 × 20
mm
Sensitivity Adjustment Range
1: 104
—
Peak Sensitivity Wavelength
400
430
nm
Luminous Sensitivity
Min.
30
Typ.
60
40
140
40
140
µA/lm
60
250
60
200
Blue Sensitivity Index (CS 5-58) Typ.
7.1
6.4
—
6.4
—
—
Red/White Ratio
Typ.
—
—
0.3
—
0.15
—
Radiant Sensitivity *1
Typ.
60
60
74
60
80
mA/W
Luminous Sensitivity *2 Min.
50
Typ.
600
200
400
200
400
2500
400
300
A/lm
700
Radiant Sensitivity *1 *2 Typ.
6.0 × 105
4.0 × 105
7.4 × 105
4 × 105
2.8 × 105
A/W
Dark Current *2 *3
Typ.
5
0.1
3
0.1
0.2
nA
Max.
50
2
50
0.5
1
Dark Count *3 *4
Typ.
—
—
—
30
80
s-1
Max.
—
—
—
80
200
Rise Time *2
Typ.
2.2
ns
Ripple Noise *2 *5 (peak to peak) Max.
0.5
mV
Settling Time *6
Max.
0.2
s
Operating Ambient Temperature *7
+5 to +45
°C
Storage Temperature *7
-20 to +50
°C
Weight
220
g
*1: Measured at the peak sensitivity wavelength *2: Control voltage = +1.0 V *3: After 30 minutes storage in darkness
*4: Control voltage: Plateau voltage
*5: Cable RG-174/U, Cable length 450 mm, load resistance = 1 MΩ, load capacitance = 22 pF
*6: The time required for the output to reach a stable level following a change in the control voltage from +1.0 V to +0.5 V. *7: No condensation