English
Language : 

H10330B-25_15 Datasheet, PDF (1/4 Pages) Hamamatsu Corporation – THERMOELECTRIC COOLED NIR-PMT UNIT
LOW-LIGHT-LEVEL THERMOELECTRIC COOLED
MEASUREMENT NIR-PMT UNIT
IN THE NIR H10330B-25/-45/-75
Wavelength Range: 950 nm to 1200 nm / 950 nm to 1400 nm / 950 nm to 1700 nm,
TE Cooled, High Speed, Suitable for Photon Counting
OVER VIEW
The H10330B series is an NIR-PMT unit using a compact NIR-PMT
(near infrared photomultiplier tube) developed by our advanced photo-
cathode technology. The NIR-PMT is contained in a thermally insulated
sealed-off housing evacuated to a high vacuum. The internal thermo-
electric cooler eliminates the need for liquid nitrogen and cooling water.
The light input window of these unit use a condenser lens to provide a
virtually larger photosensitive area allowing easy optical coupling.
Adapters for connection to an optical fiber and monochromator are also
available as options.
APPLICATIONS
GPhotoluminescence Measurement
GSinglet Oxygen Measurement
GLIDAR
GRaman Spectroscopy Measurement
GCathodoluminescence Measurement
GFluorescence, Fluorescence Life Time Measurement
GOptical Communication Device Evaluation
Left: NIR-PMT Main Unit, Right: Controller
FEATURES
GCompact and lightweight due to vacuum
sealed-off thermal insulation technology
GHigh Sensitivity (Applicable to Photon Counting)
GFast Time Response
Rise Time: 0.9 ns, T.T.S.: 400 ps
GSimple Operation by Air Cooled TE Cooler
No Liquid Nitrogen, No Cooling Water is
Necessary
GOperable in 20 min after Switched ON
GLarge Detection Area
18 mm for Collimated Light
GHV Power Supply with Interlock Function
GOptional Adapters are Available
For Optical Fiber
For Monochromator