English
Language : 

C9913GC_15 Datasheet, PDF (1/8 Pages) Hamamatsu Corporation – Mini-spectrometers
Mini-spectrometers
TG series
C9913GC
C9914GB
For near IR, integrating optical system,
image sensor and circuit
Hamamatsu TG series mini-spectrometers C9913GC, C9914GB are polychromators integrated with optical elements and an im-
age sensor. Light to be measured is guided into the entrance port of TG series through an optical fiber and the spectrum mea-
sured with the built-in image sensor is output from the USB port to a PC for data acquisition. The C9913GC and C9914GB allow
accurate measurement with low noise by cooling the image sensor. The C9913GC and C9914GB come supplied with evaluation
software that allows setting measurement conditions, acquiring and saving data, and displaying graphs. Original measurement
software can be designed on an end-user’s side as DLL’s function specification is disclosed.
Features
High throughput due to transmission grating made of quartz
Highly accurate optical characteristics
Low noise
Compact design for easy assembly
Wavelength conversion factor*1 is recorded in internal memory
Applications
Water content measurement
Component analysis in food, agriculture fields, etc.
Process control for chemical products
*1: A conversion factor for converting the image sensor pixel number into a wavelength is recorded in the module. A calculation factor
for converting the A/D converted count into the input light level is not provided.
Optical characteristics
Parameter
TG-cooled NIR-I
C9913GC
TG-cooled NIR-II
C9914GB
Unit
Spectral response range
Spectral resolution (FWHM)*2
Wavelength reproducibility*3
900 to 1700
7 max.
-0.2 to +0.2
1100 to 2200
nm
8 max.
nm
-0.4 to +0.4
nm
Wavelength temperature dependence
Spectral stray light*2 *4
-0.02 to +0.02
-35 max.
-0.04 to +0.04
nm/°C
dB
*2: Depends on the slit opening. Values were measured with the slit listed in the table “ Structure”.
*3: Measured under constant light input conditions
*4: When monochromatic light of the following wavelengths is input, spectral stray light is defined as the ratio of the count
measured at the input wavelength, to the count measured in a region of the input wavelength ±40 nm.
C9913GC: 1300 nm, C9914GB: 1650 nm
Electrical characteristics
Parameter
C9913GC
C9914GB
Unit
A/D conversion
16
bit
Integration time
5 to 10000
5 to 1000
ms
Interface
USB 1.1
-
Current consumption of USB bus power
250 max.
mA
www.hamamatsu.com
1