English
Language : 

C6386-01_15 Datasheet, PDF (1/2 Pages) Hamamatsu Corporation – Light-to-voltage conversion amplifier with optical fiber
MODULE
Photosensor amplifier
C6386-01
Light-to-voltage conversion amplifier with optical fiber
Features
l Easy handling
Built-in photodiode allows easy detection of light just by
connecting to a voltmeter.
l Optical fiber light input
Measures light at a narrow detection point. Separating the
amplifier from the detection point allows measurement in
unusual environments and achieves low noise.
l Three sensitivity ranges
Applications
l Optical power meters
l Laser monitors
l Precision photometry
l Various light detection (sparks inside equipment, etc.)
Range C o nversio n im pe dan ce (V /A ) C u t-off fre qu en cy (M H z)
H
105
1
M
104
3
L
103
10
l High-speed response
s Absolute maximum ratings
Parameter
Symbol
Condition
Supply voltage
Vs
Current consumption
Is Dark state
Output short-circuit time
-
Operating temperature
Topr No condensation
Optical fiber usage temperature
- No condensation
Minimum bending radius for
optical fiber
-
Value
Unit
±18
V
±7
mA
Continuous
-
0 to +40
°C
0 to +60
°C
R15
mm
s Recommended operating range
Parameter
Symbol
Supply voltage
Vs
Condition
Min.
Typ.
±6
-
Max.
Unit
±15
V
s Specifications (Ta=25 °C, Vs=±15 V, ±9 V using integrated dry batteries)
Parameter
Symbol
Condition
Min.
Photo sensitivity
opening diameter
-
-
Numerical aperture
-
-
Photo sensitivity *1
H
λ=830 nm
-
M
S Measured when light is input
-
L
to the optical fiber end.
-
Output resistance
-
-
Output voltage amplitude
VFS
Vs=±15 V, RL=2 kΩ
Vs=±9 V, RL=2 kΩ
+10
+5
Output noise
Vn Dark state, all bandwidth
-
Frequency bandwidth *2
H
DC
M
f -3 dB
DC
L
DC
Battery lifetime
-
With dry batteries
dark state, RL > 10 kΩ
-
W eight
- Including batteries
-
*1: Refer to spectral response
*2: Refer to frequency response
Typ.
φ2.0
0.56
30
3
0.3
200
-
-
-
1
3
10
90
460
Max.
-
-
-
-
-
-
-
-
10
-
-
-
-
-
Unit
mm
-
mV/µW
Ω
V
mVp-p
MHz
hr
g
1