C13366_15 Datasheet, PDF (1/6 Pages) Hamamatsu Corporation – Optical measurement modules for low-level-light detection, Analog output
MPPC® modules
C13366 series
(GA type)
Optical measurement modules for low-level-light detection,
Analog output
The C13366 series (GA type) are optical measurement modules capable of detecting low level light. These modules consist of
a thermoelectrically cooled MPPC, a signal amplifier circuit, a high-voltage power supply circuit, and a temperature control cir-
cuit. The photosensitive area is available in two sizes of 1.3 × 1.3 mm and 3 × 3 mm, and the signal output is analog. Mod-
ules operate just by connecting them to an external power supply (±5 V).
Features
Built-in TE-cooled MPPC
[MPPC for precision measurement (new product)]
High sensitivity in the short wavelength range
Low noise equivalent power
Built-in temperature control function
Analog output
Applications
Low-level-light measurement
Flow cytometry
Fluorescence measurement
Analytical instrument
Structure
Parameter
Symbol
C13366-1350GA
C13366-3050GA
Unit
Internal MPPC
-
TE-cooled MPPC
-
Effective photosensitive area
-
1.3 × 1.3
3×3
mm
Pixel pitch
-
50
μm
Number of pixels
-
667
3600
-
Absolute maximum ratings
Parameter
Symbol
Condition
Value
Unit
Supply voltage
Vs
±6
V
Operating temperature
Topr No dew condensation*1
-10 to +40
°C
Storage temperature
Tstg No dew condensation*1
-20 to +70
°C
*1: When there is a temperature difference between a product and the surrounding area in high humidity environment, dew condensation
may occur on the product surface. Dew condensation on the product may cause deterioration in characteristics and reliability.
Note: Exceeding the absolute maximum ratings even momentarily may cause a drop in product quality. Always be sure to use the
product within the absolute maximum ratings.
Electrical and optical characteristics (Typ. Ta=25 °C, λ=λp, Vs=±5 V, unless otherwise noted)
Parameter
Spectral response range
Peak sensitivity wavelength
Element temperature
(setting temperature)
Photoelectric sensitivity
Cutoff frequency
High band
Low band
Rise time
Noise equivalent power
Minimum detection limit
Maximum output voltage
Symbol
λ
λp
Condition
C13366-1350GA
Min.
Typ.
Max.
C13366-3050GA
Min.
Typ.
Max.
Unit
270 to 900
270 to 900
nm
-
500
-
-
500
-
nm
Td
-
-20
-
-
-20
-
°C
-
0.7 × 109 1.0 × 109 1.3 × 109 0.7 × 109 1.0 × 109 1.3 × 109 V/W
fc -3 dB, sine wave
3
4
DC
-
3
4
-
MHz
DC
-
tr 10 % to 90%, 1.p.e.
-
5
-
-
9
-
ns
NEP Dark state
-
0.1
0.2
-
0.15
0.3 fW/Hz1/2
- Dark state
-
0.25
0.5
-
0.35
0.7 pW rms
-
-
4.7
-
-
4.7
-
V
www.hamamatsu.com