English
Language : 

C13365_15 Datasheet, PDF (1/6 Pages) Hamamatsu Corporation – Optical measurement modules for low-level-light detection, Analog output
MPPC® modules
C13365 series
Optical measurement modules for low-level-light detection,
Analog output
The C13365 series are optical measurement modules capable of detecting low level light. These modules consist of an
MPPC, a signal amplifier circuit, a high-voltage power supply circuit, and a temperature compensation circuit. The photosen-
sitive area is available in two sizes of 1.3 × 1.3 mm and 3 × 3 mm, and the signal output is analog. Modules operate just
by connecting them to an external power supply (±5 V).
Features
Built-in MPPC (new product) for precision measurement
High sensitivity in the short wavelength range
Low noise equivalent power
Built-in temperature compensation circuit
Compact and lightweight
Analog output
Applications
Flow cytometry
Low-level-light measurement
Fluorescence measurement
Analytical instrument
Structure
Parameter
Symbol
C13365-1350SA
C13365-3050SA
Unit
Effective photosensitive area
-
1.3 × 1.3
3×3
mm
Pixel pitch
-
50
μm
Number of pixels
-
667
3600
-
Absolute maximum ratings
Parameter
Symbol
Condition
Value
Unit
Supply voltage
Vs
±6
V
Operating temperature
Topr No dew condensation*1
-20 to +60
°C
Storage temperature
Tstg No dew condensation*1
-20 to +80
°C
*1: When there is a temperature difference between a product and the surrounding area in high humidity environment, dew condensation
may occur on the product surface. Dew condensation on the product may cause deterioration in characteristics and reliability.
Note: Exceeding the absolute maximum ratings even momentarily may cause a drop in product quality. Always be sure to use the
product within the absolute maximum ratings.
Electrical and optical characteristics (Typ. Ta=25 °C, λ=λp, Vs=±5 V, unless otherwise noted)
Parameter
Spectral response range
Peak sensitivity wavelength
Temperature stability of
output voltage
Photoelectric sensitivity
Cutoff
High band
frequency Low band
Rise time
Noise equivalent power
Minimum detection limit
Maximum output voltage
Symbol
λ
λp
-
-
fc
tr
NEP
-
-
Condition
C13365-1350SA
Min.
Typ.
Max.
270 to 900
-
500
-
C13365-3050SA
Min.
Typ.
Max.
270 to 900
-
500
-
Ta=25 ± 10 °C
-
-
±5
-
-
±5
0.7 × 109 1.0 × 109 1.3 × 109 0.7 × 109 1.0 × 109 1.3 × 109
-3 dB, sine wave 3.5
5
DC
-
3.5
5
DC
-
10% to 90%, 1 p.e. -
5
-
-
9
-
Dark state
-
0.5
1.0
-
1.2
2.0
Dark state
-
1
2
-
2.7
4.5
-
4.7
-
-
4.7
-
Unit
nm
nm
%
V/W
MHz
-
ns
fW/Hz1/2
pW rms
V
www.hamamatsu.com