English
Language : 

C12666MA Datasheet, PDF (1/11 Pages) Hamamatsu Corporation – Micro-spectrometer
Micro-spectrometer
C12666MA
Finger-tip size, ultra-compact spectrometer head
integrating MEMS and image sensor technologies
The C12666MA is an ultra-compact (Finger-tip size) spectrometer head developed based on our MEMS and image sensor
technologies. The adoption of a newly designed optical system has achieved a remarkably small size, less than half the vol-
ume of the previous mini-spectrometer MS series (C10988MA-01). In addition, the employment of hermetic packaging has
improved humidity resistance.
This product is suitable for integration into a variety of devices, such as integration into printers and hand-held color
monitoring devices that require color management. It is also suitable for applications that collaborate with portable
devices, such as smartphones and tablets.
Features
Applications
Finger-tip size: 20.1 × 12.5 × 10.1 mm
Weight: 5 g
Color monitoring for printers and printing machines
Testers for lights and LEDs
Spectral response range: 340 to 780 nm
Color adjustment of various large size displays
Spectral resolution: 15 nm max.
Hermetic package: High reliability against humidity
Installation into mobile measurement equipment
Wavelength conversion factor*1 is listed on final inspection sheet
Water quality control monitors and other environment
measuring instruments
Measuring instruments that use portable devices
such as smartphones and tablets
*1: A conversion factor for converting the image sensor pixel number into a wavelength. A calculation factor for converting the A/D
converted count into the input light level is not provided.
Optical characteristics
Parameter
Value
Spectral response range
340 to 780
Spectral resolution (FWHM)
15 max.
Wavelength reproducibility*2
-0.5 to +0.5
Wavelength temperature dependence
Spectral stray light*3
-0.1 to +0.1
-25
*2: Measured under constant light input conditions
*3: Spectral stray light = 10 × log (Tl/Th)
Th: count measured when light at a certain wavelength is input
Tl: count measured at a wavelength 40 nm longer or shorter than the input light wavelength
Unit
nm
nm
nm
nm/°C
dB
Electrical characteristics
Parameter
Min.
Typ.
Max.
Unit
Supply voltage
4.75
5
5.25
V
Power consumption
-
30
-
mW
Video rate
Output impedance
0.25
-
-
150 *4
200
kHz
-
Ω
*4: An increase in the current consumption at the video output terminal also increases the chip temperature and so causes the dark
current to rise. To avoid this, connect a buffer amplifier for impedance conversion to the video output terminal so that the current
flow is minimized. As the buffer amplifier, use a JFET or CMOS input operational amplifier of optical input impedance.
www.hamamatsu.com
1